ID |
Date |
Author |
Type |
Category |
Subject |
10802
|
Tue Dec 16 00:20:06 2014 |
diego | Update | Optical Levers | BS & PRM OL realignment | [Rana, Diego]
We manually realigned the BS and PRM optical levers on the optical table. |
10801
|
Mon Dec 15 22:45:59 2014 |
Jenne | Update | Electronics | Yend QPD modified |
[Jenne, Rana, Diego]
We did some test on the modified QPD board for the Yend; we saw some weird oscillations at high frequencies, so we went and check more closely directly within the rack. The oscillations disappear when the cable from the QPD is disconnected, so it seems something is happening within the board itself; however, looking closely at the board with an oscilloscope in several locations, with the QPD cable connected or disconnected, there is nothing strange and definitely nothing changing if the cable is connected or not. In the plots there are the usual channels we monitor, and the 64kHz original channels before they are downsampled.
Overall it doesn't seem being a huge factor, as the RMS shows at high frequencies, maybe it could be some random noise coming up, but anyway this will be investigated further in the future. |
Attachment 1: QPD_Ytrans_oscillating_15Dec2014.pdf
|
|
Attachment 2: QPD_IOPchannels_Ytrans_oscillating_15Dec2014.pdf
|
|
10800
|
Mon Dec 15 22:40:09 2014 |
rana | Summary | PSL | PMC restored | Found that the PMC gain has been set to 5.3 dB instead of 10 dB since 9 AM this morning, with no elog entry.

I also re-aligned the beam into the PMC to minimize the reflection. It was almost all in pitch. |
10799
|
Mon Dec 15 22:30:50 2014 |
Jenne | Update | Electronics | Yend QPD modified | Details later - empty entry for a reply.
Short story - Yend is now same as Xend filters-wise and lack of gain sliders -wise. Both ends have 13.7k resistors around the AD620 to give them gains of ~4.5.
Xend seems fine.
Yend seems not fine. Even the dark noise spectrum sees giganto peaks. See Diego's elog 10801 for details on this investigation. |
10798
|
Mon Dec 15 16:27:57 2014 |
diego | Update | Computer Scripts / Programs | Status of the new nodus |
Quote: |
Nodus (solaris) is dead, long live Nodus (ubuntu).
Diego and I are smoothing out the Kinks as they appear, but the ELOG is running smoothly on our new machine.
SVN is working, but your checkouts may complain because they expect https, and we haven't turned SSL on yet...
|
[Diego, EricQ]
SSL, https and backups are now working too!
A backup of nodus's configuration (with some explaining) will be done soon. |
10797
|
Mon Dec 15 12:53:13 2014 |
ericq | Update | Computer Scripts / Programs | Status of the new nodus | Nodus (solaris) is dead, long live Nodus (ubuntu).
Diego and I are smoothing out the Kinks as they appear, but the ELOG is running smoothly on our new machine.
SVN is working, but your checkouts may complain because they expect https, and we haven't turned SSL on yet... |
10796
|
Sat Dec 13 14:26:36 2014 |
ericq | Update | LSC | Mismatched gains on ETMY Transmon QPD | Yesterday, we were seeing anomalously high low frequency RIN in the y-arm (rms of 4% or so). I swung by the lab briefly to check this out. Turns out, despite TRY of 1.0, there was reasonable misalignment. ASS with the excitation lowered by a factor of two, and overall gain at 0.5 or so aligned things to TRY=1.2, and the RIN is back down to ~0.5% I reset the Thorlabs FM to make the power = 1.0
I then went to center the transmitted beam on the transmon QPD. Looking at the quadrant counts as I moved the beam around, things looked odd, and I poked around a little...
I strongly suspect that we have significantly mismatched gains for the different quadrants on the ETMY QPD.
Reasoning: With the y-arm POY locked, I used a lens to focus down the TRY beam, to illuminate the quadrants individually. Quadrants 2 and 3 would go up to 3 counts, while 1 and 4 would go up to 0.3 and 0.6, respectively. (These counts are in some arbitrary units that were set by setting the sum to 1.0 when pitch and yaw claimed to be centered, but mismatched gains makes that meaningless.)
I haven't looked more deeply into where the mismatch is occurring. The four individual whitening gain sliders did affect the signals, so the sliders don't seem sticky, however I didn't check the actual change in gains. Will the latest round of whitening board modifications help this?
Hopefully, once this is resolved, the DC transmission signals will be much more reliable when locking... |
10795
|
Sat Dec 13 00:35:11 2014 |
rana | Update | Electronics | Xend QPD whitening board modified |
16 bit. There aren't any 14-bit ADCs anywhere in LIGO. The aLIGO suspensions have 18-bit DACs.
The Y-End gains seem reasonable to me. I think that we only use TRX/Y as error signals once we have arm powers of >5 so we should consider if the SNR is good enough at that point; i.e. what would be the actual arm motion if we are limited only by the dark noise?
I seem to remember that the estimate for the ultimate arm power is ~200, considering that we have such high losses in the arms. |
10794
|
Fri Dec 12 19:54:21 2014 |
Jenne | Update | Electronics | Xend QPD whitening board modified | Okay, I have finished modifying the Xend QPD whitening board, although I will likely need to change the gain on Monday.
Rather than following my plan in elog 10782, I removed the AD602's entirely, and just use the AD620's as the amplifiers. We don't need remotely adjustable gains, and the AD620s are a less noisy part.
I set the gain to be 30dB using a 1.65k resistor for R_G, which turns out to be too high. After I installed the board and realized that my counts were much higher than they used to be, I realized that what we had been calling +30dB was in fact +13.2dB. ( I am assuming that the ADC for the gain sliders were putting out a maximum of +10V. The AD620 used to have a 1/10 voltage divider at the input, and an overall gain of 1, so the output of the AD620 was 100mV. This goes into pin 16 of the AD602, which has gain of 32*V_set + 10. Which gives 32*0.1+10=13.2dB. Ooops. We've been lying to ourselves. )
Anyhow, before I made the gain realization, I was happily going along, setting the AD620s' gains all to 30dB. I also copied Koji's modification from April of this year, and permanently enabled the whitening filters.
Here is the schematic of what ended up happening. The red modifications were already in place, and the greens are what I did today.

You can see the "before" picture in my elog Wednesday, elog 10774. Here is an "after" photo:

Here is a spectrum comparing the dark noise of the Xend QPD after modification to the current Yend QPD (which is still using the AD602 as the main instrumentation amplifier). I have given the Yend data an extra 16.8dB to make things match.

And, here is a set of spectra comparing both ends, dark noise versus single arm lock. While I'll have to sacrifice a lot of it, there's oodles more SNR in the Xend now. The Yend data still has the "gain fixing" extra 16.8dB.

The Xend quadrant input counts (before the de-whitening filters) now go up to peak values of about 1,000 at single arm lock. If (optimistically) the we got full power recycling and the arms got to powers of 300, that would mean we would have 300,000 counts, which is obviously way more than we actually have ADC range for. Currently, the Yend quadrant input counts go as high as 50, which with arm powers of 300 would give 15,000 counts. I think I need to bring the Xend gain down to about the level of the Yend, so that we don't saturate at full arm powers. I can't remember right now - are the ends 14-bit or 16-bit ADCs? If they're 16-bit, then I can set the gain somewhere between the current X and Y values.
Finally, I added a section of the 40m's DCC document tree for the QPD whitening: E1400473, with a page for each end. Xend = D1400414, Yend = D1400415. |
10793
|
Fri Dec 12 19:38:49 2014 |
diego | Update | Computer Scripts / Programs | Status of the new nodus |
Quote: |
[Diego, Steve]
We ran a Cat 6+ Ethernet cable from the 1X7 rack (where the new nodus is located) to the fast GC switch in the control room rack; now I will learn how to setup the 'outside world' network, iptables, and the like.
I remind that the current hardware/software status is posted in elog 10697 ; if additions or corrections are needed, let me know.
After I check a couple of things, we can use the new nodus (which is currently known in the martian network as rosalba) as a local test to see that everything is working. After that (and, mostly, after I'll have the network working), we will sync the data from the old nodus to the new one and make the switch.
|
[Diego, EricQ]
Update: work is almost completed; the old nodus is still online, as I don't feel confident to make the switch and leave it on its own for the weekend. However, the new nodus is online with the IP address 131.215.114.87, so everyone can check that everything works. From my tests I can say that:
After everything will be in place, I will save every reasonably important configuration file of nodus into the svn.
I remind that every change made while accessing the 131.215.114.87 machine will be purged during the sync&switch
|
10792
|
Fri Dec 12 15:19:04 2014 |
manasa | Summary | General | Dec 12 - PSL table |
Quote: |
Unfortunately the order placed for beam samplers last week did not go through. These will be used at the X and Y end tables to dump the unwanted light appropriately. Since they will not be here until Tuesday, I revised the timeline for FOL related activities accordingly.
|
I was working on the PSL table today.
Since the rejected 1064nm light after the SHG crystal is not easily reachable to measure beam widths close to the waist, I put a lens f=300mm and measured the beam size around its focus. I used this data and redesigned the telescope using 'a la mode'.
I used a beam splitter to attenuate the beam directed towards the fiber. The reflected beam from BS has been dumped (I need to find a better beam dump than what is being used right now.
I have only ~200uW at the input of the fiber coupler after the BS and 86uW at the output of the fiber (43% coupling).
I moved the GTRY DC photodiode and the lens in front of it to make space for the fiber coupler mount.
The layout on the PSL table right now is as shown below.
I have also put the fiber chassis inside the PSL enclosure on the rack. I moved the coherent spectrum analyser controller that is not being used to make space on the rack.

|
Attachment 2: PSLfiberChassis.png
|
|
10791
|
Fri Dec 12 14:38:39 2014 |
manasa | Summary | General | Frequency Offset Locking - To Do List (Revised) | Unfortunately the order placed for beam samplers last week did not go through. These will be used at the X and Y end tables to dump the unwanted light appropriately. Since they will not be here until Tuesday, I revised the timeline for FOL related activities accordingly. |
Attachment 1: FOLtodolist.pdf
|
|
10790
|
Fri Dec 12 10:38:22 2014 |
Steve | Update | PEM | leaky roof | The first real rain of this year finds only one leak at the 40m |
Attachment 1: leakyroof.jpg
|
|
10789
|
Fri Dec 12 04:33:49 2014 |
Jenne | Update | ASC | ASS retuned | [Rana, Jenne]
We decided that tonight was the night for ASS tuning.
We started from choosing new frequencies, by looking at the transmission and the servo control signals spectra to find areas that weren't too full of peaks. We chose to be above the OpLev UGF by at least a factor of ~2, so our lowest frequency is about 18Hz. This way, even if the oplevs are retuned, or the gains are increased, the ASS should still function.
We set the peak heights for the lowest frequency of each arm to have good SNR, and then calculated what the amplitude of the higher frequencies ought to be, such that the mirrors are moving about the same amount in all directions.
We re-did the low pass filters, and eliminated the band pass filters in the demodulation part of the servo. The band passes aren't strictly necessary, as long as you have adequate lowpassing, so we have turned them off, which gives us the freedom to change excitation frequencies at will. We modified the lowpass filter so that we had more attenuation at 2Hz, since we spaced our excitation frequencies at least ~2.5 Hz apart.
The same lowpass filter is in every single demodulator filter bank (I's and Q's, for both length and transmission demodulation). We are getting the gain hierarchy just by setting the servo gains appropriately.
We ran ezcaservos to set the demodulation phase of each lockin, to minimize the Q-phase signal.
We then tuned up the gains of the servos. Rana did the Y arm, but for the X arm I tried to find the gains where the servos went unstable, and then reduced the gain by a factor of 2. The Xarm is having trouble getting good alignment if you start with something less than about 0.7, so there is room for improvement.
Rana wrote a little shell script that will save the burt snapshot, if the gains need adjusting and they should be re-saved.
The scripts have been modified (just with the new oscillator amplitudes - everything else is in the burt snapshots), so you should be able to run the start from nothing and the start from frozen scripts for both arms. However, please watch them just in case, to make sure they don't run away. |
10788
|
Fri Dec 12 02:30:25 2014 |
Jenne | Summary | General | PSL table optical layout |
Quote: |
I missed to elog this earlier. I have temporarily removed the DC photodiode for GTRY to install the fiber holder on the PSL table. So GTRY will not be seeing anything right now.
|
After some confusion, I discovered this a few hours ago. |
10787
|
Thu Dec 11 23:34:06 2014 |
manasa | Summary | General | PSL table optical layout |
Quote: |
I assembled the telescope to couple PSL light into the fiber. The maximum coupling that I could obtain was 10mW out of 65mW (~15%).
I was expecting to achieve 80-90% coupling from my design estimates. It makes me wonder if the beam waist measurements made by Harry during summer were correct in the first place. I would like to go back and check the beam waist at the PSL table.
Also, we need a pair of 8m (~25 feet) long SMA cables to carry the RF signal from the beat PD on the PSL table to frequency counter module on the IOO rack.
Steve says that we had a spool of SMA cable and it was borrowed by someone a few months ago. Any updates on either who is holding it or if it has been used up already would help.
The X end slow computer was down this morning. So I used only the Y arm ALS to record the noise level for reference. DTT data for ALSY out of loop noise before opening PSL enclosure is saved in /users/manasa/data/141211/ALSYoutLoop.xml
|
I missed to elog this earlier. I have temporarily removed the DC photodiode for GTRY to install the fiber holder on the PSL table. So GTRY will not be seeing anything right now.
|
10786
|
Thu Dec 11 22:44:23 2014 |
rana | Update | LSC | QPD screens | All of the QPDX matrix fields had a missing underscore in them. So I committed all of the c1asc screens to the SVN (because no one but me and Jamie seems to be able to remember to do this).
Then I did find/replace on the QPDY screen and saved it over the QPDX screen and committed the new thing to SVN as well. Values are now accessible. |
10785
|
Thu Dec 11 18:12:46 2014 |
ericq | Update | LSC | (Fixed) Y end whitening board |
Quote: |
Gain mystery
- It was not sure how the whitening gains have been given.
- The corresponding database entry was found in /cvs/cds/caltech/target/c1auxey/ETMYaux.db as
grecord(ao,"C1:ASC-QPDY_S1WhiteGain")
grecord(ao,"C1:ASC-QPDY_S2WhiteGain")
grecord(ao,"C1:ASC-QPDY_S3WhiteGain")
grecord(ao,"C1:ASC-QPDY_S4WhiteGain")
- The gains for S2-S4 were set to be 30. However, C1:ASC-QPDY_S1WhiteGain was set to be 8.62068.
And it was not writable.
- After some investigation, it was found that the database was wrong. The DAC channel was changed from S100 to S0.
The corrected entry is shown here.
grecord(ao,"C1:ASC-QPDY_S1WhiteGain")
{
field(DESC,"Whitening gain for QPDY Seg 1")
field(DTYP,"VMIVME-4116")
field(OUT,"#C0 S0 @")
field(PREC,"1")
field(EGUF,"42")
field(EGUL,"-22")
field(EGU,"dB")
field(LINR,"LINEAR")
field(DRVH,"30")
field(DRVL,"-10")
field(HOPR,"30")
field(LOPR,"-10")
}
- Once c1auxey was rebooted, the S1 whitening gain became writable. Now all of the channels were set to be +30dB (max).
|
This exact situation was happening at ETMX. I did the exact same change to the database, now I can read and write all four gain segments. |
10784
|
Thu Dec 11 18:00:43 2014 |
ericq | Update | Computer Scripts / Programs | MC error signal monitoring | The other day, I hooked up the agilent analyzer to OUT2 of the MC board, which is currently set to output the MC refl error signal. I've written a GPIB-based program that continuously polls the analyzer, and plots the live spectrum, an exponentially weighted running mean, and the first measured spectrum.
The intended use case is to see if the FSS or MC loops are going crazy when we're locking. Sometimes the GPIB interface hangs/loses its connection, and the script needs a restart.
The script lives in scripts/MC/MCerrmon
|
10783
|
Thu Dec 11 17:45:54 2014 |
ericq | Update | Computer Scripts / Programs | Lockloss plotting | With some advice from Jamie, I've gotten the lock loss plotting script that is used at LHO working on our machines. The other night, I modified the ALSwatch.py script to log lockloss times. Tying it together, I've written a small wrapper script that grabs the last time from the lockloss log, and plots it.
It is: scripts/LSC/LocklossData/lastlock.sh
Jamie's going to make an adjustment to the pydv codebase that will let me implement the auto y-scaling that we like. We also will need to get a feel for the right timing window, once we see what kind of delay in the ALSwatch script is typical.
Here's an example of the output, with the window of [-10,+2] seconds from the logged GPS time:

|
10782
|
Thu Dec 11 16:42:12 2014 |
Jenne | Update | Electronics | Xend QPD whitening board plan | Here is a little PDF of what I plan to do to both of the transmission QPD whitening boards later today. The idea is to take away the remote gain slider inputs, and force the gains to always be at +30dB.
The red and blue notes are from Koji's elog 9854, and the green are my plans for today.
I will cut the traces from the gain slider inputs, and pull the negative input of the AD620 to ground. The positive input will be connected to the +5 voltage line, with a divider so that the positive input to the AD620 is about 666mV.
The AD602 will be maxed out at +30dB with anything over 625mV.

Unless there are objections, I will start these modifications in an hour or so. I will also make the Xarm whitening always-on, just like Koji has already done for the Yend.
EDIT, JCD, 12Dec2014: These are not the modifications that were made. Please see ____ for actual modifications. |
10781
|
Thu Dec 11 15:40:04 2014 |
Steve | Update | safety | safety training |
Katherine Dooley has received 40m specific basic safety training in the 40m lab |
10780
|
Thu Dec 11 12:50:12 2014 |
manasa | Summary | General | PSL table optical layout | I assembled the telescope to couple PSL light into the fiber. The maximum coupling that I could obtain was 10mW out of 65mW (~15%).
I was expecting to achieve 80-90% coupling from my design estimates. It makes me wonder if the beam waist measurements made by Harry during summer were correct in the first place. I would like to go back and check the beam waist at the PSL table.
Also, we need a pair of 8m (~25 feet) long SMA cables to carry the RF signal from the beat PD on the PSL table to frequency counter module on the IOO rack.
Steve says that we had a spool of SMA cable and it was borrowed by someone a few months ago. Any updates on either who is holding it or if it has been used up already would help.
The X end slow computer was down this morning. So I used only the Y arm ALS to record the noise level for reference. DTT data for ALSY out of loop noise before opening PSL enclosure is saved in /users/manasa/data/141211/ALSYoutLoop.xml
|
10779
|
Thu Dec 11 12:39:31 2014 |
diego | Update | Computer Scripts / Programs | Frequency Offset Locking scripts status | I finished the polishing in the scripts/FOL directory, this is the current status and this post replaces my two previous posts on the subject:
- the Raspberry Pi operates locally: everything is in its /opt/FOL directory, which is a mirror of /opt/rtcds/caltech/c1/scripts/FOL/ ; some backup/sync scripts should be set up, tell me what kind (sync direction, place to call the script from, etc..) is recommended and I'll set it up;
- the /opt/FOL directory contains:
- ADC_interface : Akhil's ADC interface software and dependencies;
- akhilTestCodes : Akhil's work directory with his programs and data;
- backup : two zip files with a full backup of the FOL stuff for both chiara and domenica at 2014/12/08, before my work on the directory;
- fcreadoutApp : the EPICS app compiled on domenica. I didn't modify anything in particular here, as I don't know much about EPICS Apps; I'm not even sure if it is used by now, as I launch EPICS manually by just giving him a .db file (see below).
- armFC* : it is the single program that constantly fetches data for the channels: it takes as arguments the RPi device (/dev/hidraw0 for the X arm, and /dev/hidraw1 for the Y arm) and the value to write into the frequency counter (0x3 for initialization and 0x2 for actual use); hence, there is no more need for recompilation!
- epics_channels.py : this is the new version of the old codetorun.py script; it handles the initialization and the availability of the two beatnote channels;
- fcreadout / freqCountIOC : these are the binaries of the EPICS apps that I found on chiara/domenica; they are not used as of now, but could be useful;
- fcreadout.db : it is the database file that is loaded by EPICS to handle the channels;
- FOLPID.pl : the Perl PID controller; it is still the old version, we will work on this one later on (see Manasa's schedule at elog 10760 for info)
- Domenica's environment:
- as I said, everything runs locally from /opt/FOL;
- in particular, I added in /etc/inittab two lines that launch EPICS and the python script for the channels; respawn is supported so these processes should always be available. For this to happen, DO NOT MOVE armFC, epics_channels.py and fcreadout.db from the /opt/FOL directory on domenica!
- I added a udev rule in /etc/udev/rules.d/98-hidraw-permissions.rules to let the controls user access the /dev/hidraw* devices without having to sudo all the time;
- I updated the ~/.bashrc and /opt/epics/epics-user-env.sh files to fix syntax errors and add some aliases we usually use.
|
10778
|
Thu Dec 11 10:08:10 2014 |
manasa | Update | General | IFO update | Status of IFO:
1. The X end slow computer is down. ETMX sliders and buttons on the ETMX suspension screens have gone white. I have disabled the ETMX oplev because it is largely misaligned in yaw. I am not poking it and leaving it as is for the time being.
2. The Y arm green alignment had drifted and GTRY was down to 0.15 . I tweaked the alignment using the last two steering mirrors and brought GTRY to 0.7 which gives a beat note of -14dBm. |
10777
|
Thu Dec 11 09:11:18 2014 |
manasa | Update | PSL | PSL FSS Slow actuator | I am not sure if people have been noticing it lately; but the slow actuator on the PSL FSS has been railing up quite often these days. I found it at >0.8 and as high as 1.5 on certain occasions before resetting it to nominal zero.
It could be because the PMC alignment needs to be tweaked. The night crew should consider doing this before starting to lock. |
10776
|
Wed Dec 10 21:05:56 2014 |
Kate | Update | SEI | Guralp briefly powered down | Kate & Jenne
About 2:30 this afternoon, we briefly powered off the Guralp (C1:PEM-SEIS_GUR1_{X,Y,Z}) in order to better align it with the other seismometers along its marked N/S direction. It had been visibly off by a few degrees. |
10775
|
Wed Dec 10 16:12:29 2014 |
manasa | Summary | General | Dec 10 - PSL table |
Quote: |
Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours.
|
I was working around the PSL table today.
I wanted to modify the telescope that couples PSL light into the fiber; now that I have the translation stages for the lenses. I could not finish it as the locking work started earlier than usual this afternoon. I measured the out of loop noise for ALS error signals before I opened the PSL enclosure. X and Y beat notes were at -18dBm at 49.3MHz and -29.56dBm at 62.2MHz for this measurement. DTT data can be found in /users/manasa/data/141210/ALSoutLoop.xml; so there is reference to go back to in case of any damage done due to the work on the PSL table.
Also, I received the front and back panels for the Fiber chassis and put it together. Find photos (front panel and inside) of chassis in attachment. This will go inside the PSL enclosure tomorrow.

|
10774
|
Wed Dec 10 15:05:32 2014 |
Jenne | Update | Electronics | Xend QPD whitening board modified already | In April, Koji logged that he had made some changes to the Yend QPD whitening board (elog 9854). Today, I pulled the Xend board to see if it had the same modifications. The filter shapes all seem to be the same (as in, the capacitors at the output filters were removed, etc.), and the final gain is the same. I just realized that I didn't explicitly check if the whitening switches were pulled to ground to permanently turn on the whitenening, but hopefully I'll be able to see that in the photo.
I have not made any changes today (yet) to the board, so the overall gain is still accessible via EPICS. I wanted to do a quick check that we won't be saturating things at full power with the maximum gain, before I make a change.

EDIT: After comparing the photos here and in elog 9854, the X end board has the filter shape modifications that were done some time ago, but the whitening is not permanently enabled. For the Yend board, Koji added a jumper wire connecting (for example) R97 and R106 to the grounded side of C69. This jumper wire is not in place on the X qpd board.
Before I re-pull the board and modify it, I want to make sure I know what I'm going to do for the gain slider override. |
10773
|
Wed Dec 10 14:26:13 2014 |
steve | Update | PEM | air cond maintenance and particle plot |
Quote: |
AC maintenance is scheduled from 8am till 11am tomorrow morning.
|
IFO air condition maintenance will continue tomorrow morning and it should be finished by 11:30AM
We have new guys taking over this job: Sal and Chris so it takes longer. The units will be shut down for a bit.
They will not enter the IFO lab. The CES housed units will be worked on. |
Attachment 1: 05.png
|
|
10772
|
Wed Dec 10 14:22:37 2014 |
diego | Update | Computer Scripts / Programs | Status of the new nodus | [Diego, Steve]
We ran a Cat 6+ Ethernet cable from the 1X7 rack (where the new nodus is located) to the fast GC switch in the control room rack; now I will learn how to setup the 'outside world' network, iptables, and the like.
I remind that the current hardware/software status is posted in elog 10697 ; if additions or corrections are needed, let me know.
After I check a couple of things, we can use the new nodus (which is currently known in the martian network as rosalba) as a local test to see that everything is working. After that (and, mostly, after I'll have the network working), we will sync the data from the old nodus to the new one and make the switch. |
10771
|
Tue Dec 9 16:07:16 2014 |
manasa | Summary | General | Dec 9 - FC module and fiber chassis |
Quote: |
Quote: |
Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours.
|
Elaborate to do list:
1. The FC module should be mounted on the IOO rack. Domenica has to be powered up appropriately to the rack power supply.
2. The fiber chassis needs to be built. This will hold all the fiber components and will sit inside the PSL enclosure.
Fiber connectors and fiber couplers need to be installed in the chassis. Attached is the cartoon sketch of layout in the chassis.
3. User guide for FC module (work in progress)
|
1. FC module has been mounted on the IOO rack. The module gets it AC supply from the powerstrip already installed on the back side of the rack.

2. The fiber chassis has not been put together completely. We have still not received the front and back panels for the chassis; which is keeping me on hold. Diego is almost done with his housekeeping on Domenica. He will post an elog with all the details.
3. User guide for FC module (work in progress) |
10770
|
Tue Dec 9 16:06:46 2014 |
diego | Summary | General | Dec 8 - Check Frequency Counter module |
Quote: |
Quote: |
Quote: |
Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours.
|
[Diego, Manasa]
We looked into the configuration and settings that the frequency counters (FC) and Domenica (the R pi to which the FCs talk to) were left at . After poking around for a few hours, we were able to readout the FC output and see it on StripTool as well.
We have made a list of modifications that should be done on Domenica and to the readout scripts to make the FC module automated and user-friendly.
I will prepare a user manual that will go on the wiki once these changes are made.
|
I started working on the scripts/FOL directory (I did a backup before tampering around!):
- I still need to make some serious polishing in the folder, and into the Raspberry Pi itself, in order to have a clean and understandable environment;
- as of now, I created an single armFC.c program, which takes as arguments the device (/dev/hidraw0 for the X arm, and /dev/hidraw1 for the Y arm) and the value to write into the frequency counter (0x3 for initialization and 0x2 for actual use); hence, no more need for recompilation!
- I improved the codetorun.py script (and gave the fellow a proper name, epics_channels.py) which handles the initialization AND the availability of the channels;
- On the Raspberry Pi, I created two init scripts, /etc/init.d/epics_server.sh and /etc/init.d/epics_channels.sh, which start at the end of the boot process with default runlevels; the former starts the softIOc process (epics itself), while the latter executes the constantly running epics_channels.py script; as they are services, they can be started/stopped with the usual sudo /etc/init.d/NAME start|stop|restart
As a result, as soon as the Raspberry Pi completes its boot process, the two beatnote channels are immediately available.
|
Update and corrections:
- I forgot to log that I added a udev rule in /etc/udev/rules.d/98-hidraw-permissions.rules in order to let the controls user access the devices without having to sudo all the time;
- I updated the ~/.bashrc and /opt/epics/epics-euser-env.sh files to fix syntax errors and add some aliases we usually use;
- since /etc/init.d/ doesn't support automatic respawn of processes, I purged the two scripts I did yesterday and added two lines to /etc/inittab. This works just as fine (I tried a couple of reboots to verify that) and the two processes now respawn automatically even if killed (and, I assume, if they die for any other reason)
- Another thing I forgot: for the time being, during the cleanup, the Raspberry Pi works on the network share script directory. Once cleaning is done and everything is fixed, everything will run locally on the RPi, and the scripts/FOL directory on chiara will be used as backup/repository.
|
10769
|
Tue Dec 9 03:41:06 2014 |
Jenne | Update | CDS | EPICS running slow - network issue? | [Jamie, EricQ, Jenne, Diego]
This is something that we discussed late Friday afternoon, but none of us remembered to elog.
We have been noticing that EPICS seems to run pretty slowly, and in fact twice last week froze for ~2 minutes or so (elog 10756).
On Friday, we plotted several traces on StripTool, such as C1:SUS-ETMY_QPD_SUM_OUTPUT and C1:SUS-ETMY_TRY_OUTPUT and C1:LSC-TRY_OUTPUT to see if channels with (supposedly) the same signal were seeing the same sample-holding. They were. The issue seems to be pretty wide spread, over all of the fast front ends. However, if we look at EPICS channels provided by the old analog computers, they do not seem to have this issue.
So, Jamie posits that perhaps we have a network switch somewhere that is connected to all of the fast front end computers, but not the old slow machines, and that this switch is starting to fail.
My understanding is that the boys are on top of this, and are going to figure it out and fix it. |
10768
|
Tue Dec 9 03:34:52 2014 |
Jenne | Update | ASC | POP yaw razor tuning | With the re-do of the IFO alignment last week, I think that the beam was no longer about halfway on the POP22 razor blade. To fix this, I locked the PRMI on sideband, removed the razor blade, and then put it back in such that it occluded about half of the light.
I'm not entirely sure why, but when I put the razor in, POP22 went from 104(ish) to 45(ish) but POPDC went from 5200(ish) to 1600(ish). [The 'ish'es are because the PRC wasn't angularly stabilized, so there was some motion changing the power levels that leaked out to the POP port]. The ETMs were misaligned, so this should not be a carrier vs. sideband effect, since they'll both share the cavity axis defined by the ITMs and the PRM. It is possible, although I didn't check, that there is some oplev light scattered into the POP photodiode that is now blocked by the razor blade. This light would only be at DC and not the 2f frequencies. Since the signal levels for POP22 vs. POPDC didn't change with and without the table top on (and with and without room lights on), I don't think that it is an effect of ambient light getting into the diode. To check if it is oplev light I should (a) just look, and (b) try to lock the PRMI without the ITMX oplev laser being on to see if there is a difference in the POPDC signal.
Anyhow, under the assumption that the POP22 signal level is correct, I tuned up the PRCL ASC a little bit. These changes are now in the carm_cm_up script, and the carm_cm_down script resets things. Before the PRC is locked, I have FM1 and FM7 (the basic servo shape and a 40Hz lowpass) on, the gain set to zero, and the input off. After lock is acquired, the input is turned on, and the gain ramps from 0 -> 10 in 3 seconds. Then FM2 and FM6 (boosts at 1 and 3Hz) are engaged.
In the plot below, the dark blue and red curves were taken when there was no angular control on the PRC. Pink was taken last week with the old QPD yaw ASC on. Light blue is today's version of the in-loop performance of the POP22 yaw ASC loop. I didn't save the trace unfortunately, but the DC QPD saw out-of-loop improvement between about 0.8Hz - 4 Hz.
Also, has anything happened with the LSC rack in the last few weeks that might be causing lots of 60Hz noise? I saw these large lines last week, but I don't think I remember them from the past.

After I got the PRCL ASC working, I tried several iterations of locking. ETMX is still being annoying, although the last hour or so have been okay. CARM keeps getting rung up right around the transition to the sqrtInv error signal. Since CARM and DARM are kind of entangled, it took me a few iterations to figure out that it was CARM that is ringing up, and not DARM. I'm a little worried about the phase loss from the 1kHz lowpass that we turn on just before the transition to sqrtInv. I want to keep the lowpass off until after we have transitioned DARM also over to DC transmission. I tried once, but I lost lock before starting the CARM transition. Anyhow, the ETM alignment issue is annoying.
Also, Jamie, Q, Diego and I were discussing last Friday, but none of us elogged, that we think there might be something wrong with one of the Martian network switches. I'll start a separate thread about that right now, but it slows things down when you can't trust EPICS channels to be current, and I (without evidence) am a little worried that this might also affect the fast signals. |
10767
|
Tue Dec 9 00:30:27 2014 |
manasa | Summary | General | Dec 9 - Elaborate to do list |
Quote: |
Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours.
|
Elaborate to do list:
1. The FC module should be mounted on the IOO rack. Domenica has to be powered up appropriately to the rack power supply.
2. The fiber chassis needs to be built. This will hold all the fiber components and will sit inside the PSL enclosure.
Fiber connectors and fiber couplers need to be installed in the chassis. Attached is the cartoon sketch of layout in the chassis.
3. User guide for FC module (work in progress) |
Attachment 1: FOL_FiberChassis.pdf
|
|
10766
|
Mon Dec 8 20:53:51 2014 |
diego | Summary | General | Dec 8 - Check Frequency Counter module |
Quote: |
Quote: |
Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours.
|
[Diego, Manasa]
We looked into the configuration and settings that the frequency counters (FC) and Domenica (the R pi to which the FCs talk to) were left at . After poking around for a few hours, we were able to readout the FC output and see it on StripTool as well.
We have made a list of modifications that should be done on Domenica and to the readout scripts to make the FC module automated and user-friendly.
I will prepare a user manual that will go on the wiki once these changes are made.
|
I started working on the scripts/FOL directory (I did a backup before tampering around!):
- I still need to make some serious polishing in the folder, and into the Raspberry Pi itself, in order to have a clean and understandable environment;
- as of now, I created an single armFC.c program, which takes as arguments the device (/dev/hidraw0 for the X arm, and /dev/hidraw1 for the Y arm) and the value to write into the frequency counter (0x3 for initialization and 0x2 for actual use); hence, no more need for recompilation!
- I improved the codetorun.py script (and gave the fellow a proper name, epics_channels.py) which handles the initialization AND the availability of the channels;
- On the Raspberry Pi, I created two init scripts, /etc/init.d/epics_server.sh and /etc/init.d/epics_channels.sh, which start at the end of the boot process with default runlevels; the former starts the softIOc process (epics itself), while the latter executes the constantly running epics_channels.py script; as they are services, they can be started/stopped with the usual sudo /etc/init.d/NAME start|stop|restart
As a result, as soon as the Raspberry Pi completes its boot process, the two beatnote channels are immediately available.
|
10765
|
Mon Dec 8 15:54:39 2014 |
manasa | Summary | General | Dec 8 - Check Frequency Counter module |
Quote: |
Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours.
|
[Diego, Manasa]
We looked into the configuration and settings that the frequency counters (FC) and Domenica (the R pi to which the FCs talk to) were left at . After poking around for a few hours, we were able to readout the FC output and see it on StripTool as well.
We have made a list of modifications that should be done on Domenica and to the readout scripts to make the FC module automated and user-friendly.
I will prepare a user manual that will go on the wiki once these changes are made.
|
10761
|
Mon Dec 8 09:03:49 2014 |
steve | Update | PEM | air cond maintenance tomorrow morning |
AC maintenance is scheduled from 8am till 11am tomorrow morning. |
10760
|
Sun Dec 7 13:11:57 2014 |
manasa | Summary | General | Frequency Offset Locking - To Do List | Attached is the timeline for Frequency Offset Locking related activities. All activities will be done mostly in morning and early afternoon hours. |
Attachment 1: FOLtodolist.pdf
|
|
10759
|
Fri Dec 5 15:41:45 2014 |
Jenne | Update | Computer Scripts / Programs | Nodus on fast GC network | Apparently, some time ago Larry Wallace installed a new, fast ethernet switch in the old nodus rack. Q and I have just now moved nodus' GC ethernet cable over to the new switch. Dan Kozak is going to use this faster connection to make the data flow over to the cluster not so lag-y. |
10758
|
Fri Dec 5 02:44:43 2014 |
Jenne | Update | General | IFO alignment shenanigans | [Jenne, Q, Diego]
OMG, today sucked alignment-wise. Like, wow.
I think that the problem with the ASS is with the input pointing part of the system. I found that if I disable the TTs for the Yarm (iin practice, the outputs are held at zero), I could run the Yarm ASS at full gain of 1, and it would do an awesome job. The first time I did this, I by-hand optimized the TTs by running the test mass loops to make them follow the input pointing. After that, I haven't touched the TT pointing at all, and we've just been running the test mass loops for the Yarm ASS. The Xarm seems to not have this problem (or at least not as drastically), so I left it as it was, touching BS as well as ITMX and ITMY, although the gain still needs to be about 0.3.
I feel pretty good about the IFO alignment now, although it is slightly different than it has been. The transmitted arm powers are higher than they were before I changed the ASS procedure, and there seems to be a little less power fluctuation with alignment. Q points out that I don't have concrete evidence that this is a good alignment, but it feels right.
It was a significant enough change that I had to go down to the Yend to realign the green to the new arm axis. Xgreen we did with the remote PZTs. I also realigned both of the beatnotes on the PSL table.
While I was on the PSL table, I quickly touched up the PMC alignment.
The biggest problem, the one that sucked up more hours and energy than I'd like to admit, is ETMX's jumping. So frustrating. Sometimes it is time-coincident with engaging the LSC, sometimes not. I thought that it might be because there are too many violin filters, but even if I turn off all violin filters to ETMX, it jumped a few times while the cavity was locked. Sometimes it moves when the cavity is just locked and seems happy, sometimes it moves when nothing is resonating except for the green. It takes a few minutes to recover the alignment enough to lock, and then it'll jump again a few minutes later. I haven't gone down to squish the cables today, although I did it yesterday and that didn't seem to do anything.
We had a few hours of time when it wasn't jumping, so we tried a few times to lock the IFO. The last several times we have lost lock because the PRC loop rang up. We measured the loop at low-ish arm powers, but it kept losing lock at higher powers before we could measure. At least 3 times, the PRC lockloss took out CARM and DARM too.
Anyhow, it has been a long day of not accomplishing anything interesting, but hopefully the IFO will feel better tomorrow. |
10757
|
Fri Dec 5 00:52:51 2014 |
Jenne | Update | SUS | ETMX 2nd order violin | We looked at the spectra of POX and POY during IR lock, and Q saw a peak at 1285 in POX only. We're actuating on the ETMs, so it must be an ETMX violin mode, although it doesn't match the others that are in the table.
Anyhow, I added it to FM9. While I was doing that, I realized that yesterday I had forgotten to put back the 3rd order ETM violin notch, so that is also in FM9. |
10756
|
Thu Dec 4 23:45:30 2014 |
Jenne | Update | CDS | Frozen? | [Jenne, Q, Diego]
I don't know why, but everything in EPICS-land froze for a few minutes just now. It happened yesterday that I saw, but I was bad and didn't elog it.
Anyhow, the arms stayed locked (on IR) for the whole time it was frozen, so the fast things must have still been working. We didn't see anything funny going on on the frame builder, although that shouldn't have much to do with the EPICS service. The seismic rainbow on the wall went to zeros during the freeze, although the MC and PSL strip charts are still fine.
After a few minutes, while we were still trying to think of things to check, things went back to normal. We're going to just keep locking for now.... |
10755
|
Thu Dec 4 03:09:06 2014 |
Jenne | Update | LSC | Bump in Darm Plant: Lockloss after measurement | We were sitting around arm powers of 6, and that loop measurement had finished. I was about to go down to arm powers of 5ish, but we lost lock. I'm not sure why. There's some slow stuff going on in some of the servos, but nothing jumps out at me as a loop oscillation. It does however kind of look like the PRMI lost lock just before the arm powers went down? Perhaps this somehow triggered a lockloss?
The time is 1101721675.
Wide view plots:




Zooms:




We're stopping for tonight because ETMX is back to its lame-o jumping around. I went in and squished the cables, but it's still happening.
Also, the FSS PC drive has been high the last few minutes (only starting after we quit for the night). When the MC re-locks, it sounds like an ocean wave dying out as the noise goes down a little bit. But, after a few minutes, it'll get mad again and unlock the MC.
Also, also, I noticed this on Monday with Diego, but the LSC-ALS[x,y] filter module gains sometimes mysteriously get set to zero. WTF? Eric and I have both independently checked, and we cannot find a single script in the scripts directory with the string "LSC-ALS", so we aren't deliberately changing those. Does anyone know what might be going on here? |
10754
|
Thu Dec 4 02:59:05 2014 |
ericq | Update | LSC | Bump in Darm Plant | [J,Q]
After some housekeeping (ASS is wonky, alignment of X green beat was bad, tuning of demod angles, fm gains for REFL165), we were able to bring the PRFPMI up to arm powers of 8 very stably.
We were keeping an eye on the DARM OLG, to make sure the gain was correct. We then saw a bump around 120Hz. Here is the bump.

Changing CARM offset changes its amplitude. Maybe it's a DARM optical spring. It didn't occur to me until after we lost lock that we could have tweaked the DARM offset to move it around if this was the case.
Unfortunately, due to some unexplained locklosses, we weren't able to get back into a state to measure this more... which is annoying. During that stable lock, Jenne stated that PRCL and DARM noises were looking particularly good.
We may want to tweak the way we handle the transmission PD handoff; maybe we want to force the switch at a certain place in the carm_up script, so that we're not flipping back and forth during an IR handoff; I think this may have been responsible for a lock loss or two. |
10753
|
Thu Dec 4 01:24:47 2014 |
Jenne | Update | SUS | PRM volin 3rd harmonic | Earlier this afternoon, while locking PRMI, I saw a big peak at 1883.48 Hz. This comes closest to the PRM's 627.75 Hz *3, so I infer that it is the 3rd order harmonic of the PRM violin mode.
While putting this in, I noticed that my addition of ETM filters the other day (elog 10746) had gotten deleted. Koji pointed out that Foton can do this - it allows you to create and save filters that are higher than 20th order, but secretly it deletes them. I went into the filter archive and recovered the old ETM filters, and split things up. I have now totally reorganized the filters, and I have made every single optic (ETMs, ITMs, PRM, SRM, BS, MC2) all the same.
FM1 is BS 1st and 2nd harmonics, and FM6 directly below that is a generic 3rd order notch that is wide enough that it encompases 3*BS.
FM2 is the PRM 1st and 2nd order, and FM7 below it is the PRM 3rd order.
FM3 is the SRM 1st order, FM4 is the ETMs' 1st order, and FM5 is the MC2 1st and 2nd order filters.
All of these filters are triggered on if any degree of freedom is triggered. They all have a ramp time of 3 sec. We may want to consider having separate trigger options for each optic, so that we're not including the PRM notch on the ETMs, for example, and vice versa.
When all of these filters are on, according to Foton we lose 5.6 degrees of phase at 100 Hz. |
10752
|
Thu Dec 4 00:26:07 2014 |
Jenne | Update | ASC | POP yaw razor blade installed | We would like the option of feeding back the POP beam position fluctuations to the PRM to help stabilize the PRC since we don't have oplevs for PR2 and PR3. However, we cannot just use the DC QPD because that beam spot will be dominated by carrier light as we start to get power recycling.
The solution that we are trying as of today is to look at yaw information of just the RF sidebands. (Yaw is worse than pitch, although it would be nice to also control pitch). I have placed a razor blade occluding about half of the POP beam in front of the POP PD (which serves POPDC, POP22 and POP110). I also changed the ASS model so that I could use this signal to feed back to the PRM. Loop has been measured, and in-loop spectra shows some improvement versus uncontrolled.
Optical table work:
The POP beam comes out of the vacuum system and is steered around a little bit, then about 50% goes to the DC QPD. Of the remaining, some goes to the Thorlabs PD (10CF I think) and the rest goes to the POP camera. For the bit that goes to the Thorlabs PD, there is a lens to get the beam to fit on the tiny diode.
There was very little space between the steering mirror that picks off the light for this PD, and the lens - not enough to put the razor blade in. The beam after the lens is so small that it's much easier to occlude only half of the beam in the area before the lens. (Since we don't know what gouy phase we're at, so we don't know where the ideal spot for the razor is, I claim that this is a reasonable place to start.)
I swapped out the old 50mm lens and put in a 35mm lens a little closer to the PD, which gave me just enough room to squeeze in the razor blade. This change meant that I had to realign the beam onto the PD, and also that the demod phase angles for POP22 and POP110 needed to be checked. To align the beam, before placing the razor blade, I got the beam close enough that I was seeing flashes in POPDC large enough to use for a PRMI carrier trigger. The PRMI carrier was a little annoying to lock. After some effort, I could only get it to hold for several seconds at a time. Rather than going down a deep hole, I just used that to roughly set the POP22 demod phase (I -phase maximally negative when locked on carrier, Q-phase close to zero). Then I was able to lock the PRMI sideband by drastically reducing the trigger threshold levels. With the nice stable sideband-locked PRMI I was able to center the beam on the PD.
After that, I introduced the razor blade until both POPDC and POP22 power levels decreased by about half.
Now, the POP22 threshold levels are set to up=10, down=1 for both MICH and PRCL, DoF triggers and FM triggers.
ASS model work:
POP22 I and POP110 I were already going to the ASS model (where ASC lives) for the PRCL ASS dither readbacks. So, I just had to include them in the ASC block, and increased the size of the ASC input matrix. Now you can select either POP QPD pit, POP QPD yaw, POP221 or POP110I to go to either PRCL yaw, PRCL pit, CARM yaw or CARM pit.
Compiled, installed and restarted the ASS model.
Engaging the servo:
I took reference spectra of POP QPD yaw and POP 22, before any control was applied. The shapes looked quite similar, but the overall level of POP22 was smaller by a factor of ~200. I also took a reference spectra of the POP QPD in-loop signal using the old ASC loop situation.
Q looked at Foton for me, and said that with the boost on, the UGF needed to be around 9 or 10 Hz, which ended up meaning a servo gain of +2.5 (the old POP QPD yaw gain was -0.063). We determined that we didn't know why there was a high-Q 50Hz notch in the servo, and why there is not a high frequency rolloff, so right now the servo only uses FM1 (0:2000), FM6 (boost at 1Hz and 3Hz) and FM7 (BLP40).
The in-loop residual isn't quite as good with POP22 as for the QPD, but it's not bad.
Here's the loop:

And here's the error spectra. Pink solid and light blue solid are the reference traces without control. Pink dashed is the QPD in-loop. Red and blue solid are the QPD and POP22 when POP22 is used as the error signal. You can definitely see that the boosts in FM6 have a region of low gain around 1.5Hz. I'm not so sure why that wasn't a problem with the QPD, but we should consider making it a total 1-3Hz bandpass rather than a series of low-Q bumps. Also, even though the POP22 UGF was set to 9 Hz, we're not seeing any suppression above about 4Hz, and in fact we're injecting a bit of noise between 4-20Hz, which needs to be fixed still.

|
10751
|
Wed Dec 3 21:41:12 2014 |
Jenne | Update | Computer Scripts / Programs | Matlab license updated | It seems that the old Matlab servers went down a week or so early, so I have updated the Matlab license information in
/cvs/cds/caltech/apps/linux64/matlab/licenses/
per the instructions on https://www.imss.caltech.edu/content/updating-matlab-license-file
EDIT: Q did this also for the control room iMac |
10750
|
Wed Dec 3 08:36:49 2014 |
Steve | Update | LSC | good IFO status | |
Attachment 1: goodStatus.png
|
|
|