Fri May 9 01:36:28 2014, rana, Summary, SUS, Optical Lever QPD Sum trends: they're almost all too weak
|
Mon May 12 22:42:19 2014, rana, Summary, SUS, Optical Lever QPD Sum trends: they're almost all too weak 
|
Mon May 12 22:52:59 2014, Jenne, Summary, SUS, Optical Lever Filters are all different
|
Tue May 13 13:27:58 2014, Steve, Summary, SUS, Optical Lever Servos setting table
|
Tue May 13 17:03:05 2014, Steve, Summary, SUS, ETMX oplev laser replaced
|
Tue May 13 17:31:32 2014, Jenne, Summary, SUS, ETMX oplev laser replaced: New oplev gains set 
|
Tue May 13 22:55:57 2014, Jenne, Summary, SUS, ETMX oplev: cleanup
|
Wed May 14 10:04:06 2014, Steve, Summary, SUS, oplev laser and temperature
|
Tue May 20 09:12:25 2014, Steve, Summary, SUS, ETMX oplev qpd board  
|
Tue May 20 15:54:39 2014, rana, Summary, SUS, ETMX oplev qpd board
|
|
Message ID: 9973
Entry time: Tue May 20 09:12:25 2014
In reply to: 9952
Reply to this: 9975
|
Author: |
Steve |
Type: |
Summary |
Category: |
SUS |
Subject: |
ETMX oplev qpd board |
|
|
Quote: |
ETMX oplev qpd gain has to be increased.
|
Atm3, Oplev sum read 12,000 counts when the qpd was disconnected ?
Dark qpd was zero and normal He/Ne incident on qpd was 1,730 counts. |
|
|
|
|
|