Message ID: 9891     Entry time: Thu May 1 13:03:34 2014     In reply to: 9888     Reply to this: 9892
Author: Jenne 
Type: Update 
Category: LSC 
Subject: MC board pulled for AC coupling 

Quote:

To fix this, I am going to put a 6.8uF cap in series with R30 in the MC board, which is part of the crossbar switch where the IN1 and IN2 get summed.  This should AC-couple the output of the IN2 slider before the summing node.

 MC board is out, so don't be surprised that the MC isn't locking.