Message ID: 9851     Entry time: Thu Apr 24 22:39:14 2014     In reply to: 9835     Reply to this: 9875
Author: rana 
Type: Configuration 
Category: General 
Subject: netgpibdata is working again now 

Quote:

Not working again. I tried the commands in Koji's elog as well as the ones from my notes, but the AG 4395 hooked up to the yellow box named crocetta doesn't work. It gets to the stage of opening the data files and hangs. I tried it with many variations on pianosa and rossa. Also tried power cycling the analyzer and the Prologix and the bridge.

While hooked up to the MC error point I set the modulation frequency from 137 to 133 Hz to minimize the 3 MHz peak as usual.

Started working tonight. Don't know if anyone did anything to the Martian network or not, so its a mystery...

I also modified the script and SVN'd it. It now correctly takes your plot wants and adjust the linear/log of the axes accordingly

./SPAG4395A.py -i crocetta -A -v 4 --att=0 --start=1kHz --end=10MHz --bw=1kHz --plot --semiy

and also saves the plot as out.pdf

In the attached image I show the MC board TP1A output. The two peaks around 3.7 Mhz are the sideband beats we speak of. The lower one is proportional to the MC length/frequency mistmatch.

Attachment 1: out.pdf  20 kB  | Show | Show all