Jenne made some suggestions for some plots that would be useful on our CARM offset reduction adventures, so I made some with my MIST model.
First, here's a plot showing the transfer function of CARM to TRX, with logarithmically spaced offsets out to 3nm. While not a control signal, it shows us where the optical plant resonance stuff is happening. The peaks in this TF correspond to peaks in REFL11, REFL55, AS11, etc., as in the close-to-resonance TFs in ELOG 9785.

[more to come, had a MATLAB issue]
|