40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Apr 11 01:02:07 2014, Jenne, Update, LSC, CARM and DARM both on IR signals!!!!!!!!! onlyIRspectra.pdf
    Reply  Fri Apr 11 02:09:31 2014, Jenne, Update, LSC, CARM and DARM both on IR signals!!!!!!!!! IRlocks.pdf
       Reply  Fri Apr 11 11:58:24 2014, Jenne, Update, LSC, CARM and DARM both on IR signals!!!!!!!!! DARM_IN1_getsSmaller.pngDARM_CARM_TRX_TRY.pngCARMtransition_DARMtransition.pngDARMin_DARMout.pngCARM_DARM_in_out.png
          Reply  Fri Apr 11 12:21:27 2014, Koji, Update, LSC, CARM and DARM both on IR signals!!!!!!!!! 
             Reply  Fri Apr 11 12:32:33 2014, ericq, Update, LSC, CARM and DARM both on IR signals!!!!!!!!! 
                Reply  Fri Apr 11 14:57:53 2014, steve, Update, LSC, congratulation whenSheisHappy3h.pngmc4hArms3hrs.png
                   Reply  Fri Apr 11 16:04:31 2014, Koji, Update, LSC, congratulation 
          Reply  Sun Apr 13 13:03:34 2014, rana, Update, LSC,  
    Reply  Fri Apr 11 10:30:48 2014, jamie, Update, LSC, CARM and DARM both on IR signals!!!!!!!!! 
Message ID: 9796     Entry time: Fri Apr 11 01:02:07 2014     Reply to this: 9797   9798
Author: Jenne 
Type: Update 
Category: LSC 
Subject: CARM and DARM both on IR signals!!!!!!!!! 

[EricQ, Jenne]

We're still working, but I'm really excited, so here's our news:  We are currently holding the IFO on all IR signalsNo green, no ALS is being used at all!!!!  

PRCL and MICH in REFL33, CARM on 1/sqrt(trans), DARM on AS55 Q.

CARM actuating on MC2, DARM actuating +ETMY, -ETMX. 

CARM offset is 1.9 counts, TRX averages about .1 counts. At this offset, we are able to transition CARM from ALS to DC Transmission signals and DARM from ALS to AS55Q. 