Quote: |
I'm checking the status from home.
P1 is 8e-4 torr
nodus did not feel the power outage (is it APS supported?)
linux1 booted automatically
c1ioo booted automatically.
c1sus, c1lsc, c1iscex, c1iscey need manual power button push.
|
9:11pm closed PSL shutter, turned Innolight 2W laser on,
turned 3 IFO air cond on,
CC1 5.1e-5 torr, V1 is closed, Maglev has failed, valve configuration is " Vacuum Normal " with V1 & VM1 closed, RGA not running, c1vac1 and c1vac2 were saved by UPS,
(Maglev is not connected to the UPS because it is running on 220V)
reset & started Maglev.........I can not open V1 without the 40mars running...........
Rossa is the only computer running in the control room,
Nodus and Linux1 was saved by UPS,
turned on IR lasers at the ends, green shutters are closed
It is safe to leave the lab as is.
|