40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m elog  Not logged in ELOG logo
Entry  Mon Mar 4 16:46:31 2013, yuta, Update, SUS, oplev calibration for PRM PRM_PIT.pngPRM_YAW.png
    Reply  Tue Mar 5 17:09:30 2013, yuta, Update, SUS, oplev calibration for ITMY ITMY_PIT.pngITMY_YAW.png
       Reply  Tue Apr 2 15:58:40 2013, Jenne, Update, SUS, oplev calibration for ITMX, BS opl_itmx.pngopl_bs.png
          Reply  Tue Apr 2 16:13:10 2013, rana, Update, SUS, oplev calibration for ITMX, BS 
             Reply  Tue Apr 2 16:55:34 2013, Jenne, Update, SUS, oplev calibration online for ITMs, BS, PRM comparison_opl.pdf
                Reply  Tue Apr 2 18:19:30 2013, Jenne, Update, SUS, BS, PRM oplev servos improved PRM_servo_matches_BS.pdfprm_bs_optical_levers_comparison.pdf
                Reply  Tue Apr 30 22:40:57 2013, Jenne, Update, SUS, oplev XY-plots reflect new calibration 
                   Reply  Sun Sep 22 22:05:10 2013, rana, Update, SUS, oplev XY-plots reflect new calibration noise.png
                      Reply  Thu Oct 3 01:06:48 2013, rana, jenne, Update, SUS, oplev XY-plots reflect new calibration etmx_ol.png
                         Reply  Thu Oct 3 09:49:49 2013, Steve, Update, SUS, ETMY oplev laser is sick 
                      Reply  Tue Nov 5 19:55:12 2013, rana, Update, SUS, oplev XY-plots reflect new calibration 
Message ID: 9351     Entry time: Tue Nov 5 19:55:12 2013     In reply to: 9152
Author: rana 
Type: Update 
Category: SUS 
Subject: oplev XY-plots reflect new calibration 

I used the same OSEM SUSPIT/YAW method as before to calibrate the ETMY optical lever signals. They were off by a factor of ~10.

ETMY Pitch     300  /  26    (old/new)     urad/counts

ETMY Yaw       300  /  31    (old/new)     urad/counts

These should be redone with the Kakeru / Ottaway arm cavity power technique if we want to get better than ~30% accuracy.