Message ID: 9351
Entry time: Tue Nov 5 19:55:12 2013
In reply to: 9152
|
Author: |
rana |
Type: |
Update |
Category: |
SUS |
Subject: |
oplev XY-plots reflect new calibration |
|
|
I used the same OSEM SUSPIT/YAW method as before to calibrate the ETMY optical lever signals. They were off by a factor of ~10.
ETMY Pitch 300 / 26 (old/new) urad/counts
ETMY Yaw 300 / 31 (old/new) urad/counts
These should be redone with the Kakeru / Ottaway arm cavity power technique if we want to get better than ~30% accuracy.
|