40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 9321     Entry time: Thu Oct 31 00:06:45 2013
Author: masayuki 
Type: Update 
Category: Computer Scripts / Programs 
Subject: cds fft,tf,offsets 

I wrote the scripts for cdsutils.

1. fft

usage: fft.py [-h] [-c N_CYCL] [-a N_AVG] freq channel [channel ...]

Measures the frequency content of one or more NDS channels
at the specified frequecy.
    The measurement results are  magnitude, phase, real part imaginary part, and the standard deviation of the real and imaginary parts.

positional arguments:
  freq        measurement frequency
  channel     list of measurement channel

optional arguments:
  -h, --help  show this help message and exit
  -c N_CYCL   define number of cycles. Default is 10
  -a N_AVG    define number of average. Default is 100

2. tf

usage: tf.py [-h] [-c N_CYCL] [-a N_AVG] [--dB] freq input output

Measures the transfer funtion of one NDS channels pair  at the specified frequecy.
    The measurement results are the coherence, magnitude, phase, real part, imaginary part, and the standard deviation of the real and imaginary parts.

positional arguments:
  freq        measurement frequency
  input       list of measurement input channel
  output      list of measurement output channel

optional arguments:
  -h, --help  show this help message and exit
  -c N_CYCL   define number of cycles. Default is 10
  -a N_AVG    define number of average. Default is 100
  --dB        print the amplitude in dB

 

3.offsets

usage: offset.py [-h] [-t SECONDS] filterbank [filterbank ...]

Zero the offset of one or more filterbank. Take average of OUT16 channel, and put (-1 * average / gain) into offset.  After change offsets, it will turn on the offset.
    example usage) offset -d 25 C1:LSC-AS55_Q C1:LSC-AS55_I

positional arguments:
  filterbank  list of filterbank to zero offset

optional arguments:
  -h, --help  show this help message and exit
  -t SECONDS  define the averaging time. default value is 10

I put these scripts in /opt/rtcds/cdsutils/trunk/lib/cdsutils.
I couldn't put them into cds library, but I will left tomorrow to Japan...

So please help me Jamie or Joe !!

 

 

By the way,

I modified the LSCoffset script  (script)/LSC/LSCoffsets.py.

usage: LSCoffsets.py [-h] [-d SECONDS] [--PDlist]

 Zero the offsets of LSC PDs. During taking average, we will close the PSL and green shutter. After zeroing, it will turn on the offsets switch.
    example usgae) LSCoffset2.py -d 20    

optional arguments:
  -h, --help  show this help message and exit
  -d SECONDS  define the averaging time. default value is 10
  --PDlist    print PD list for LSC and exit

i made new directory 'offset_backup' for old offset script. I move all old offset script into there.

But unfortunately that cannot use right now, because cds offsets script is not available yet...