Subject: Seismometer status 

 It would be nice if we could use the existing seismometer cable and place a 2-terminal temperature sensor within the stainless-steel can. A device like the AD590/592 can drive current over a long cable run without pickup issues since its a current source. Inside of the seismometer breakout box we should make a circuit to scale the signal to be close to zero at 25 C and have a slope of 1 V/deg. There are example circuits in the application note - we can just make them on a piece of vector board and glue to the inside of the breakout box (where we connect to the regulated power).