Wed Sep 3 19:09:56 2008, Yoichi, DAQ, Computers, c1iovme power cycled
|
Thu Sep 4 00:38:14 2008, rana, Update, PSL, c1iovme power cycled
|
Thu Sep 4 07:29:52 2008, Yoichi, Update, PSL, c1iovme power cycled
|
|
Message ID: 917
Entry time: Wed Sep 3 19:09:56 2008
Reply to this: 918
|
Author: |
Yoichi |
Type: |
DAQ |
Category: |
Computers |
Subject: |
c1iovme power cycled |
|
|
When I tried to measure the sideband power of the FSS using the scan of the reference cavity, I noticed that the RC trans. PD signal was not
properly recorded by the frame builder.
Joe restarted c1iovme software wise. The medm screen said c1iovme is running fine, and actually some values were recorded by the FB.
Nonetheless, I couldn't see flashes of the RC when I scanned the laser frequency.
I ended up power cycling the c1iovme and run the restart script again. Now the signals recorded by c1iovme look fine.
Probably, the DAQ boards were not properly initialized only by the software reset.
I will re-try the sideband measurement tomorrow morning. |