I have modified the ASS model to also have an ASS for SRCL. The input options are POPDC, POP110, AS110. I suppose I could/should have included ASDC.
Screens are modified / made. I haven't finished setting the servo gains and oscillator amplitudes, and all that jazz yet.
Using the parameters that Koji had in elog 9116, I was able to get nice long DRMI locks (several on a ~10 minute time scale).
I tried some pseudo-ANDing for the triggers, to no avail. I was trying to have the trigger matrix row for the SRCL loop have 1*POP22 and 0.02*AS110, where the 0.02 is to scale AS110 so that it has a similar amplitude to POP22. I then set threshold levels to ~250 for up, and 100 for down (I tried several different values for the up threshold). I was watching the TRIG_MON_FAST channels for both PRCL and SRCL, and I wasn't able to get SRCL to be triggered only at the same times as PRCL using this technique. Since we can get the DRMI to lock, perhaps my AND logic for the triggers is a low priority, but I think we'll need something like that if we want real logic.
|