Entry  Thu Aug 1 12:12:35 2013, gautam, Update, CDS, New model for endtable PZTs C1ASX_GDS_TP.png
    Reply  Thu Aug 1 13:09:17 2013, gautam, Update, CDS, New model for endtable PZTs-procedure 
    Reply  Thu Aug 1 15:06:59 2013, jamie, Update, CDS, New model for endtable PZTs 
       Reply  Thu Aug 1 15:28:44 2013, gautam, Update, CDS, New model for endtable PZTs-problem solved 
    Reply  Thu Aug 1 20:58:56 2013, gautam, Update, CDS, New model for endtable PZTs-MEDM Screens setup MAIN_SCREEN.pdfMATRICES.pdfLOCKINS.pdfCONTROL_FILTERS.pdf
       Reply  Mon Aug 5 18:18:32 2013, gautam, Update, CDS, Choosing LO Amplitudes and Frequencies  GTR_Power_Spectrum.pdf
          Reply  Mon Aug 5 19:10:01 2013, Koji, Update, CDS, Choosing LO Amplitudes and Frequencies  
             Reply  Tue Aug 6 16:36:51 2013, gautam, Update, CDS, Choosing LO Amplitudes and Frequencies-revised powerSpec0806.pdf
                Reply  Wed Aug 7 23:40:49 2013, gautam, Update, CDS, X-End Green ASS - A first update time-series.pdf
                   Reply  Sat Aug 10 05:53:51 2013, gautam, Update, CDS, X-End Green ASS - Roundup New_ASX_MEDM_MAIN.pdf
                      Reply  Mon Aug 12 12:57:59 2013, Jenne, Update, CDS, X-End Green ASS - Roundup 
                         Reply  Mon Aug 12 13:30:33 2013, Jamie, Update, CDS, X-End Green ASS - Roundup 
Message ID: 8952     Entry time: Thu Aug 1 15:28:44 2013     In reply to: 8951
Author: gautam 
Type: Update 
Category: CDS 
Subject: New model for endtable PZTs-problem solved 

Quote:

 

I don't know what's going on here (why the channels are white), and I don't yet have a suggestion of where to look to fix it but...

Is there a reason that you're making a new model for this?  You could just use and existing model at c1iscex, like the c1scx, and put your stuff in a top-names block.  Then you wouldn't have to worry about all of the issues with adding and integrating a new model.

Koji just fixed this.

It seems that the new model's channels were not automatically added to the master file in the framebuilder (/opt/rtcds/caltech/c1/target/master). Adding the following two lines to the master file fixed the problem;

/opt/rtcds/caltech/c1/chans/daq/C1ASX.ini

/opt/rtcds/caltech/c1/target/gds/param/tpchn_c1asx.par

The box is now green. It looks like C1ASX.ini is created automatically in /opt/rtcds/caltech/c1/chans but the master file needs to be manually edited. The channels are now showing up on dataviewer etc. I have updated the information on the wiki page.


 

 

 

 

 