Message ID: 881
Entry time: Mon Aug 25 15:50:18 2008
Reply to this: 1167
|
Author: |
rana |
Type: |
Summary |
Category: |
PEM |
Subject: |
Ranger SS-1 |
|
|
The manual for the Ranger SS-1 seismometer can be found on line here:
ftp://ftp.kmi.com/pub/software_manuals/300190/300190nc.pdf
and now in our 40m PEM Wiki page:
Ranger_SS-1
To calibrate it, we use the formula from the manual:
R_x
G_L = G_0 * ------------ = 149 +/- 3 V/(m/s)
R_x + R_c
where
G_0 = 340 V/(m/s) (generator constant)
R_x = 4300 Ohms (external damping resistor in Pomona box)
R_c = 5500 Ohms (internal coil resistance)
Then we have a gain of 200 in the SR560 so that gets us to ~30000 V/(m/s).
And then there's a DAQ conversion factor of the usual 2^16 cts / 4 V.
so the calibration constant is
G = 488 counts / (micron/sec)
in the ~1-50 Hz band |