Message ID: 8336     Entry time: Mon Mar 25 15:01:33 2013     In reply to: 8282     Reply to this: 8337
Author: Manasa 
Type: Update 
Category: Locking 
Subject: TT1 drift 

Quote:

2. TT1 drifting in pitch (Bistable)
During the arm alignment routine for spot centering, we observed that TRY dropped (from TRY = 0.9 until the arm lost lock) every 40minutes or so. The arm was relocked by moving TT1 in pitch. The (locking - unlocking due to drift - relocking) cycle was monitored and we observed that it was bistable i.e. if TT1 was moved up in pitch (0.2 on the slider) to relock for the first time ; the next time it lost lock, TT1 had to be moved down by nearly the same distance to relock the arm.
Moving TT2 or the testmasses did not help with relocking the arms; so TT1 seems to be the one causing all th trouble atleast for today.

Update

TT1 was moved in pitch to restore flashing in the arms on Wednesday (Mar 20) so that it doesn't drift too far off making it difficult to lock again. Since then, the arms have been flashing without any drifting and TTs have remained untouched. The hysteresis has disappeared.