Category: Computers 
Subject: c1susvme1 minor problems 
Around 11 am c1susvme1 start having issues. Namely C1:SUS-PRM_FE_SYNC was railing at some large value like 16384 (2^14). I presume this means the computer was running catastophically late.

I turned off the BS and ITM watch dogs (the PRM was already off), tried hitting reset and sshing in, and running startup, but this didn't help. I then turned off the c1susvme2 associated watch dogs (MC1-3, SRM) and went out to do a hard reboot by switching the crate power off. c1susvme2 came back up fine, was restarted and associated watch dogs turned back on. However, c1susvme1 came back up without mounting /cvs/cds/.

As a test, I replaced the ethernet connection with a CAT6 cable to the Prosafe switch in 1Y6, and then ran reboot on c1susvme1. When it came back up, it had mounted properly, and I was able to run the ./startup.cmd file. At this point it seems to be happy. The new cable is in the trays coming in from the top of the 1Y4 and 1Y6 and approriately labeled.

Edit: Apparently ITMX and ITMY became excited after the reboot (perhaps I turned the watchdogs back on too early? Although that was after the DAQ light was listed as green for c1susvme). Steve noticed this when the alarms went off again (I had turned them off after the reboot seemed successful), and he damped them. Interestingly, the BS remained unexcited.