Sat Feb 23 07:14:58 2013, yuta, Update, LSC, can't lock Y arm
|
Sat Feb 23 14:52:03 2013, Jenne, Update, LSC, ETMYT camera back to normal
|
Sat Feb 23 16:54:24 2013, Jenne, Update, LSC, can't lock Y arm
|
Sat Feb 23 17:14:59 2013, yuta, Update, LSC, can't lock Y arm
|
|
Message ID: 8143
Entry time: Sat Feb 23 07:14:58 2013
Reply to this: 8145
8149
|
Author: |
yuta |
Type: |
Update |
Category: |
LSC |
Subject: |
can't lock Y arm |
|
|
I tried to align and lock Y arm for the first time after pumping.
But I couldn't lock Y arm for more than ~1 sec. Why?
What I did:
1. Centered IPANG/IPPOS using input TT1/TT2.
2. Restored ITMY/ETMY slider values when it was aligned before pumping. I saw tiny flashes in TRY PD at this point.
3. Replaced Ygreen REFL camera with ETMYT camera to see transmitted beam mode.
4. Used TT1/TT2 and ITMY/ETMY to get ~ 0.4 peak in normalized TRY PD output (C1:LSC_TRY_OUT).
5. Centered POY beam on POY11 PD.
6. Changed I/Q mixing angle (C1:LSC-POY11_PHASE_R) from -61 deg to -16 deg to get good PDH signal in I_ERR (attached).
7. Ran LSCoffsets script (now on LSC_OVERVIEW screen) to adjust PD offsets.
8. Tried to lock Yarm with different gains, but failed. When lock is acquired, TRY fluctuates ~50 % and unlocks suddenly.
What I found:
1. There was some OFFSETs left turned on in suspension screens. Don't leave them on!! They change alignment of the optics. I will leave it on until we complete Yarm alignment.
2. C1:SUS-(ITMY|ETMY)_ASC(PIT|YAW) was kept oscillating the optic since Dec 17, 2013. I think this is from interrupted ASS script. Your script should restore everything when interrupted!
Next:
- Beamspot on ITMY looks off-centered. Maybe A2L is causing unstable lock?
- Maybe F2A is causing unstable lock?
- More alignment?
- FSS related? crontab related? |
|
|