We tried to lock half-PRC tonight, but we couldn't. Why?? I could lock yesterday.
It locks for ~ 1 sec, but it beam spot motion freaks out mainly in yaw.
I tried to balance PRM coils, but oplev beam was clipped by MMT1......
What I did:
1. Found elog #5392 and found F2P_LOCKIN.py
2. Modified F2P_LOCKIN.py because LOCKIN channel names are some how changed like this;
LOCKIN1_I -> LOCKIN1_DEMOD_I
LOCKIN1_Q -> LOCKIN1_DEMOD_Q
LOCKIN1_SIG -> LOCKIN1_DEMOD_SIG
3. Running
/opt/rtcds/caltech/c1/scripts/SUS/F2P_LOCKIN.py -o PRM
should adjust (UL|UR|LR|LL)COIL_GAINs by putting some gain imbalance and shaking the mirror in different frequencies. It uses LOCKIN to OL(PIT|YAW).
4. Since there was no PRM oplev beam coming out from the vacuum, I quickly looked into BS-PRM chamber. Oplev beam was clipped by MMT1. If I adjust PRM slider values to avoid clipping, the beam will be clipped by mirrors on oplev table. What happened to the PRM oplev?
5. I also made bunch of /opt/rtcds/userapps/trunk/sus/c1/medm/templates/SUS_SINGLE_LOCKIN(1|2)_DEMOD_(I|Q|SIG).adl because there were missing screens.
Next:
We need to restore the PRM oplev and balance the coils. See, also, elog #7679 |