40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Feb 6 20:19:52 2013, Manasa, Update, Optics, G&H and LaserOptik mirrors 
    Reply  Thu Feb 7 14:10:25 2013, Manasa, Update, Optics, LaserOptik - AR Reflectivity - Bad data SN6_R.pngSN6_R.png
       Reply  Fri Feb 8 21:14:52 2013, Manasa, Update, Optics, G&H - AR Reflectivity  pr2.png27_1044419003.bmp
          Reply  Fri Feb 8 22:49:31 2013, Koji, Update, Optics, G&H - AR Reflectivity  
             Reply  Fri Feb 8 23:04:40 2013, Manasa, Update, Optics, G&H - AR Reflectivity  
          Reply  Sat Feb 9 19:34:34 2013, rana, Update, Optics, G&H - AR Reflectivity  
          Reply  Mon Feb 11 19:55:47 2013, Manasa, Update, Optics, G&H - AR Reflectivity  
             Reply  Wed Feb 13 09:28:56 2013, Steve, Update, Optics, G&H - HR plots 
Message ID: 8023     Entry time: Thu Feb 7 14:10:25 2013     In reply to: 8018     Reply to this: 8045
Author: Manasa 
Type: Update 
Category: Optics 
Subject: LaserOptik - AR Reflectivity - Bad data 

Reflectivity of AR surface of LaserOptik (SN6)

 SN6_R.png

The first step measurements of R for AR surface. I am not convinced with the data....because the power meter is a lame detector for this measurement.

I'm repeating the measurements again with PDs. But below is the log R plot for AR surface.

R percentage

6000ppm @ 42 deg
3560ppm @ 44 deg
7880ppm @ 46 deg
4690ppm @ 48 deg

 

SN6_R.png

ELOG V3.1.3-