40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Wed Jan 16 18:08:06 2013, Jenne, Update, Locking, Expected PRC gains
|
Wed Jan 16 20:27:16 2013, rana, Update, Locking, Expected PRC gains
|
Thu Jan 17 00:17:31 2013, Jenne, Update, Locking, Expected PRC gains
|
Wed Jan 23 18:16:11 2013, Manasa, Update, General, Laseroptik mirror - SN6
|
Wed Jan 23 20:24:05 2013, Koji, Update, General, Laseroptik mirror - SN6
|
Wed Jan 23 22:02:25 2013, Manasa, Update, General, Laseroptik mirror - SN6
|
Thu Jan 24 15:16:50 2013, Manasa, Update, General, Laseroptik mirror - SN6
|
Thu Jan 24 16:23:24 2013, Koji, Update, General, Laseroptik mirror - SN6
|
Tue Jan 29 14:20:02 2013, Koji, Update, General, Finer rotation stage for optics characterization
|
Thu Jan 24 16:34:56 2013, rana, Update, General, Laseroptik mirror - SN6
|
Fri Jan 25 08:10:42 2013, Steve, Update, General, Laseroptik mirrors
|
Mon Jan 28 19:07:45 2013, Manasa, Update, General, SN6 Laseroptik mirror - Tranmittance measurements 
|
|
Message ID: 7947
Entry time: Mon Jan 28 19:07:45 2013
In reply to: 7944
|
Author: |
Manasa |
Type: |
Update |
Category: |
General |
Subject: |
SN6 Laseroptik mirror - Tranmittance measurements |
|
|
I repeated the measurements using NPRO instead of Crystalaser. I am attaching optical layouts for these measurements for future reference.
Lesson learnt : Do not use Crystalaser for transmission measurements and always separate the transmitted main beam from other beams that result from the wedged surface of the mirror.

Measurements match specs provided by Laseroptik
p-polarization
T percentage = 0.10% 42 deg
0.092% 44 deg
0.086% 46 deg
Minimum transmittance = 0.081% 52deg
s-polarization
T percentage = 0.048% 42 deg
0.047% 44 deg
0.047% 46 deg
Minimum transmittance = 0.047% 46 deg

|