40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Wed Jan 16 18:08:06 2013, Jenne, Update, Locking, Expected PRC gains
|
Wed Jan 16 20:27:16 2013, rana, Update, Locking, Expected PRC gains
|
Thu Jan 17 00:17:31 2013, Jenne, Update, Locking, Expected PRC gains
|
Wed Jan 23 18:16:11 2013, Manasa, Update, General, Laseroptik mirror - SN6
|
Wed Jan 23 20:24:05 2013, Koji, Update, General, Laseroptik mirror - SN6
|
Wed Jan 23 22:02:25 2013, Manasa, Update, General, Laseroptik mirror - SN6
|
Thu Jan 24 15:16:50 2013, Manasa, Update, General, Laseroptik mirror - SN6
|
Thu Jan 24 16:23:24 2013, Koji, Update, General, Laseroptik mirror - SN6
|
Tue Jan 29 14:20:02 2013, Koji, Update, General, Finer rotation stage for optics characterization
|
Thu Jan 24 16:34:56 2013, rana, Update, General, Laseroptik mirror - SN6
|
Fri Jan 25 08:10:42 2013, Steve, Update, General, Laseroptik mirrors
|
Mon Jan 28 19:07:45 2013, Manasa, Update, General, SN6 Laseroptik mirror - Tranmittance measurements 
|
|
Message ID: 7935
Entry time: Wed Jan 23 22:02:25 2013
In reply to: 7932
Reply to this: 7940
|
Author: |
Manasa |
Type: |
Update |
Category: |
General |
Subject: |
Laseroptik mirror - SN6 |
|
|
Quote: |
Got confused (even after I talked with Manasa).
The plot shows the number ~0.01 or less at 45deg. But the number is the text does not match with the plot.
Please use the logarithmic scale for the vertical axis.
And more points between 35 to 50 deg please (like ~1deg spacing)
Don't we have the data sheet from the coater? Can we request it?
|
I corrected the typo in the text...however, I agree the plot was lame ...Will get the data sheet made tomorrow! |