Entry  Thu Jan 3 20:34:43 2013, Riju, Update, , Photodiode transimpedance 
    Reply  Fri Jan 4 13:23:10 2013, rana, Update, Electronics, Photodiode transimpedance 
       Reply  Tue Jan 8 14:07:04 2013, Riju, Update, Electronics, Photodiode transimpedance 
          Reply  Tue Jan 8 15:28:41 2013, Riju, Update, Electronics, Photodiode transimpedance analyzernoiseV.pdf
Message ID: 7875     Entry time: Fri Jan 4 13:23:10 2013     In reply to: 7874     Reply to this: 7881
Author: rana 
Type: Update 
Category: Electronics 
Subject: Photodiode transimpedance 

You have to correct this transimpedance ratio by correcting for the different levels of DC photocurrent in the two devices.

For the dark noise, you must always include a trace showing the noise of the measurements device (i.e. the analyzer noise must be less than the dark PD noise) with the same input attenuation setting.