40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Oct 24 11:49:07 2012, Jenne, Update, Alignment, Game plan for the day 
    Reply  Wed Oct 24 14:19:01 2012, Jenne, Update, Alignment, Phase map summary of LaserOptik mirrors 
       Reply  Wed Oct 24 18:42:39 2012, Manasa, Update, Computer Scripts / Programs, Phase map summary of LaserOptik mirrors 12x
          Reply  Wed Oct 24 22:48:46 2012, janosch, Update, Computer Scripts / Programs, Phase map summary of LaserOptik mirrors 
             Reply  Thu Oct 25 02:01:15 2012, Koji, Update, Computer Scripts / Programs, Phase map summary of LaserOptik mirrors 
Message ID: 7606     Entry time: Wed Oct 24 11:49:07 2012     Reply to this: 7608
Author: Jenne 
Type: Update 
Category: Alignment 
Subject: Game plan for the day 

Jamie has arranged for phase map measurements this afternoon, so I will take the 6 dichroic LaserOptik optics over to Downs at 1:15 this afternoon.

Team Jamie+Nic will lead the effort to clamp down dog clamps as placement markers for all 4 in-vac passive TTs, and then pull all 4 TTs out of the chambers.  They plus Den will move the TTs to the Cleanroom, and will start to install the new pitch alignment hardware. 

When I return with the optics, we will install them in the TTs and re-balance them.  Then we can put them back in the chambers and get back to work on alignment.  

After we re-install the TTs, we will need to check the leveling of all 3 corner tables, just to be sure.

ELOG V3.1.3-