40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Sep 18 20:32:42 2012, Manasa, Configuration, PSL, AOM installation 
    Reply  Wed Sep 19 11:39:37 2012, Manasa, Configuration, PSL, AOM installation ALC_25.png
       Reply  Wed Sep 19 15:41:27 2012, Manasa, Configuration, PSL, AOM installation 
          Reply  Wed Sep 19 23:17:25 2012, rana, Configuration, PSL, AOM installation 
             Reply  Thu Sep 20 01:29:04 2012, Manasa, Configuration, PSL, AOM installation TEK00000.PNG
                Reply  Fri Sep 21 12:12:56 2012, Manasa, Configuration, PSL, AOM installation PSL.png
                   Reply  Tue Oct 2 16:15:22 2012, Manasa, Configuration, PSL, AOM installation psl_aom.png
                      Reply  Wed Oct 3 16:52:16 2012, Manasa, Configuration, PSL, AOM installation 
                         Reply  Wed Oct 3 23:36:54 2012, Koji, Configuration, PSL, AOM installation PMCTRANS.png
                            Reply  Thu Oct 4 17:54:59 2012, Manasa, Configuration, PSL, AOM installation 
                               Reply  Thu Oct 4 18:48:04 2012, janosch, Configuration, PSL, AOM installation 
                                  Reply  Thu Oct 4 20:57:43 2012, Manasa, Configuration, PSL, AOM installation 
                                     Reply  Thu Oct 4 22:16:28 2012, Koji, Configuration, PSL, AOM installation 
                                        Reply  Fri Oct 5 18:08:17 2012, Manasa, Configuration, PSL, AOM installation 
Message ID: 7464     Entry time: Tue Oct 2 16:15:22 2012     In reply to: 7425     Reply to this: 7471
Author: Manasa 
Type: Configuration 
Category: PSL 
Subject: AOM installation 

Quote:

    {Jan, Manasa}

We installed the AOM driver back on the PSL table this morning. To calibrate the AOM RF output we connected a 1V dc to the modulation input of the driver and we are convinced with the setup.

Before we direct the rf signal to the AOM, in order to check its diffraction efficiency, we would like to setup an rf PD at the AOM output. We think we have place for a filter and PD after the AOM (replacing a beam dump) and would like to confirm the position before we actually install them. The layout is the picture below showing sweet spots for the new pd to sit. If you think it may disturb the system in any way, let us know!

 

The rf PD and filter have been installed at the earlier proposed spot on the PSL table.  

psl_aom.png