40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Jul 22 03:03:22 2008, John, Rob, Update, PSL, FSS open loop transfer function 
    Reply  Tue Jul 22 13:16:09 2008, John, Rob, Update, PSL, FSS open loop transfer function 
Message ID: 715     Entry time: Tue Jul 22 13:16:09 2008     In reply to: 711
Author: John, Rob 
Type: Update 
Category: PSL 
Subject: FSS open loop transfer function 

Quote:
With the common gain slider maxed out the unity gain frequency is 58kHz.

The reference cavity refl diode appears to be okay. RF OUT/ TEST IN transfer function was normal.
There is a ~220mV offset in the RF out. We removed this using a coupler - no change. We also checked the
diode->FSS cable.

Tomorrow I'll take a closer look at the board.


Should note that the UGF of 58kHz was measured with the test cable (from RFPD to board), so the demod phase was presumably sub-optimal.