Message ID: 7145     Entry time: Fri Aug 10 16:39:44 2012
Author: Eric 
Type: Summary 
Category: Locking 
Subject: Michelson Locking 

I'm working on locking the Michelson now in order to put an excitation on one of the input test masses and measure the resulting error signal at the anti-symmetric port. I aligned the beams from ITMX and ITMY by looking at the AS camera with the video screens, but the fringes were not destructively interfering. Jenne advised that I look at the gain on the MICH servo filter modules in the LSC screen. We flipped the sign on the gain (it was 0.120 and it is now -0.120) and the fringes destructively interfered as desired after this change.

For purposes of documentation, I locked the YARM earlier in the morning before moving on to the Michelson. The purpose of this was to put another excitation on C1:SUS-ETMY_LSC_EXC and then measure the error signal on C1:LSC-POY11_I_ERR.