40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Apr 30 21:05:32 2012, Jenne, Update, SUS, ETMY oplev power supply dead 
    Reply  Mon Apr 30 21:24:43 2012, Koji, Update, SUS, ETMY oplev power supply dead 
       Reply  Mon Apr 30 22:41:34 2012, Jenne, Update, SUS, ETMY oplev power supply dead 
          Reply  Fri May 18 11:19:17 2012, steve, Update, SUS, ETMY oplev laser is dead 
             Reply  Tue May 22 08:03:01 2012, steve, Update, SUS, ETMY oplev gets new laser 
                Reply  Tue May 29 15:55:22 2012, steve, Update, SUS, ETMY oplev spot size reduced ETMYoplev.jpgIMG_1243.JPG
                   Reply  Tue May 29 21:34:08 2012, Jenne, Update, SUS, ETMY oplev spot size reduced 
                   Reply  Wed Jun 6 14:36:58 2012, steve, Update, SUS, ETMY oplev work finished 
                   Reply  Mon Oct 1 17:03:01 2012, steve, Update, SUS, ETMY oplev relayed IMG_1673oplevETMY.jpg
Message ID: 6706     Entry time: Tue May 29 15:55:22 2012     In reply to: 6656     Reply to this: 6710   6766   7458
Author: steve 
Type: Update 
Category: SUS 
Subject: ETMY oplev spot size reduced 

Quote:

Quote:

 

 The typical sign of a dying gas laser is that it glows for a few minutes only. The power supplies are fine.

Two new  JDS - Uniphase 1103P lasers ( NT64-104 )  arriving on Monday, May 21

 Yesterday I swapped in new He/Ne laser with output power 3.5 mW  The return spot on qpd is large ~6mm in diameter and 20,500 counts

The spot size reduction require similar layout as ETMX oplev.

 The oplev path is relayed and the spot size on the qpd is reduced. I still have to clean up and replace "Miki Mouse" lens holder.

There was no IP-ANG coming out of the chamber at this time!

Attachment 1: ETMYoplev.jpg  431 kB  Uploaded Mon Jun 4 10:35:31 2012  | Hide | Hide all | Show all
ETMYoplev.jpg
Attachment 2: IMG_1243.JPG  302 kB  Uploaded Mon Jun 4 10:36:35 2012  | Show | Hide all | Show all
ELOG V3.1.3-