Entry  Tue May 22 13:06:11 2012, Jenne, Update, IOO, WFS didn't turn off automatically 
    Reply  Wed May 23 21:32:15 2012, Suresh, Update, IOO, WFS didn't turn off automatically 
Message ID: 6660     Entry time: Tue May 22 13:06:11 2012     Reply to this: 6669
Author: Jenne 
Type: Update 
Category: IOO 
Subject: WFS didn't turn off automatically 

I just sat down in the control room, and discovered the PMC (and everything else) unlocked.  I relocked the PMC, but the MC wasn't coming back.  After a moment of looking around, I discovered that the WFS were on, and railing.  I ran the "turn WFS off" script, and the MC came back right away, and the WFS came on as they should.

We need to relook at the WFS script, or the MC down script, to make sure that any time the MC is unlocked, no matter why it unlocked, the WFS output is off and the filter histories are cleared.