40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 63     Entry time: Mon Nov 5 14:44:39 2007
Author: waldman 
Type: Update 
Category: OMC 
Subject: PZT response functions and De-whitening 
The PZT has two control paths: a DC coupled path with gain of 20, range of 0 to 300 V, and a pair of 1:10 whitening filters, and an AC path capacitively coupled to the PZT via a 0.1 uF cap through a 2nd order, 2 kHz high pass filter. There are two monitors for the PZT, a DC monitor which sniffs the DC directly with a gain of 0.02 and one which sniffs the dither input with a gain of 10.

There are two plots included below. The first measures the transfer function of the AC monitor / AC drive. It shows the expected 2 kHz 2d order filter and an AC gain of 100 dB, which seems a bit high but may be because of a filter I am forgetting. The high frequency rolloff is the AA and AI filters kicking in which are 3rd order butters at 10 kHz.

The second plot is the DC path. The two traces show the transfer function of DC monitor / DC drive with and with an Anti-dewhitening filter engaged in the DC drive. I fit the antidewhite using a least squares routine in matlab constrained to match 2 poles, 2 zeros, and a delay to the measured complex filter response. The resulting filter is (1.21, 0.72) : (12.61, 8.67) and the delay was f_pi = 912 Hz. The delay is a bit lower than expected for the f_pi = 3 kHz delay of the AA, AI, decimate combination, but not totally unreasonable. Without the delay, the filter is (1.3, 0.7) : (8.2, 13.2) - basically the same - so I use the results of the fit with delay. As you can see, the response of the combined digital AntiDW, analog DW path is flat to +/- 0.3 dB and +/- 3 degrees of phase.

Note the -44 dB of DC mon / DC drive is because the DC mon is calibrated in PZT Volts so the TF is PZT Volts / DAC cts. To calculate this value: there are (20 DAC V / 65536 DAC cts)* ( 20 PZT V / 1 DAC V) = -44.2 dB. Perfect!

I measured the high frequency response of the loop DC monitor / DC drive to be flat.
Attachment 1: 07110_DithertoVmonAC_sweep2-0.png  57 kB  Uploaded Wed Nov 7 12:14:52 2007  | Hide | Hide all | Show all
07110_DithertoVmonAC_sweep2-0.png
Attachment 2: 071105_LSCtoVmonDC_sweep4-0.png  67 kB  Uploaded Wed Nov 7 12:21:13 2007  | Hide | Hide all | Show all
071105_LSCtoVmonDC_sweep4-0.png
Attachment 3: 07110_DithertoVmonAC_sweep2.pdf  41 kB  Uploaded Wed Nov 7 12:21:29 2007  | Hide | Hide all | Show all
07110_DithertoVmonAC_sweep2.pdf 07110_DithertoVmonAC_sweep2.pdf
Attachment 4: 071105_LSCtoVmonDC_sweep4.pdf  22 kB  Uploaded Wed Nov 7 12:21:52 2007  | Show | Hide all | Show all
ELOG V3.1.3-