Schott, green welding glass, shade 14, 3 mm thick  was measured in the beam path of 1.2W, S polarization of 1064nm at ~1 mm diameter size as MC reflected path.

Absorption 95%, R 5% at incident angle 25-50 degrees. It looks like the perfect material for beam trap.

 