[Jenne, Zach, Frank]

Frank helped Zach and I cable up at PT-100 RTD, and make sure it worked with the Newport Model 6000 Laser Diode Controller.  We're using this rather than the Newport 3040 Temperature Controller because Dmass says the output of that isn't working.  So we're using just the temp control part of the Laser Diode controller.

The back of the controller has a 15-pin D-sub, with the following useful connections.  All others were left Not Connected. 

1 & 2 (same) - Pin 2 is one side of TEC output (we have it connected to one side of a resistive heater)

3 & 4 (same) - Pin 4 is the other side of the TEC output (connected to the other side of the resistive heater)

7 - connected to one side of PT-100 temp sensor

8 - connected to other side of PT-100 temp sensor

I used aluminum tape to attach the sensor and heater to the 40m's EOM, and we plugged in the controller.  It seems to be kind of working.  Zach figured out the GPIB output stuff, so we can talk to it remotely.

 