40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Oct 18 21:19:18 2011, rana, Configuration, IOO, WFS disabled in SUS 
    Reply  Tue Oct 18 22:47:09 2011, Suresh, Configuration, IOO, MC autolocker script edited to shutdown and restart WFS loops  
Message ID: 5688     Entry time: Tue Oct 18 21:19:18 2011     Reply to this: 5689
Author: rana 
Type: Configuration 
Category: IOO 
Subject: WFS disabled in SUS 

I found that the MC WFS had large offset control signals going to the MC SUS. Even though the input switch was off, the integrators were holding the offset.

I have disabled the ASCPIT outputs in the MC SUS. Suresh is going to fix the MC autolocker script to gracefully handle the OFF and ON and then test the script before resuming the WFS testing.

MCL data for OAF may be suspect from this morning.