I took REFL11 out from the AS table for a health check because it wasn't working properly.
The symptoms were :
- a big offset of ~ -3 V on the RF output. No RF signals.
- The DC output seemed to be okay. It's been sensitive to light.
I did a quick check and confirmed that +/- 5V were correctly supplied to the op-amps.
It looks that the last stage (MAX4107) is saturated for some reasons. Need more inspections.
At the moment the REFL11 RFPD is on the bench of the Jenne laser.
|