40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue May 3 13:22:13 2011, Leo, Update, SUS, Re: DRMI prep : suspension diagnostic 10x
    Reply  Tue May 3 15:04:20 2011, kiwamu, Update, SUS, Re: DRMI prep : suspension diagnostic 
    Reply  Wed May 4 13:57:04 2011, kiwamu, Update, SUS, mechanical resonances updated 
Message ID: 4611     Entry time: Tue May 3 13:22:13 2011     Reply to this: 4613   4626
Author: Leo 
Type: Update 
Category: SUS 
Subject: Re: DRMI prep : suspension diagnostic 

Here are the free-swinging spectra for the BS, ETMX, ETMY, ITMX, ITMY, MC1, MC2, MC3, and PRM chambers.  Kiwamu left the suspensions free for 5 hours this weekend, starting at Sat Apr 30 00:15:26 2011.

This is GPS time 988 182 941.  Quick tip: you can do local to GPS time conversions using lalapps_tconvert, which is a lot like tconvert but with special powers.  It is installed on pianosa.

$ lalapps_tconvert Sat Apr 30 00:15:26 2011

988182941

I generated these figures with the attached Python script, measure.py.

Notice that the C1:SUS-ITMX_SENSOR_UL and C1:SUS-MC3_SENSOR_UL spectra fall as 1/f.  Jenne suggested that this might indicate that there is a loose electrical connection.

Also, notice that C1:SUS-ETMY_SENSOR_LR, C1:SUS-ITMY_SENSOR_LL, and C1:SUS-PRM_SENSOR_SIDE are a lot noisier above 10 Hz.

Attachment 1: BS.png  274 kB  | Hide | Hide all | Show all
BS.png
Attachment 2: ETMX.png  289 kB  | Hide | Hide all | Show all
ETMX.png
Attachment 3: ETMY.png  311 kB  | Hide | Hide all | Show all
ETMY.png
Attachment 4: ITMX.png  301 kB  | Hide | Hide all | Show all
ITMX.png
Attachment 5: ITMY.png  274 kB  | Hide | Hide all | Show all
ITMY.png
Attachment 6: MC1.png  266 kB  | Hide | Hide all | Show all
MC1.png
Attachment 7: MC2.png  276 kB  | Hide | Hide all | Show all
MC2.png
Attachment 8: MC3.png  307 kB  | Hide | Hide all | Show all
MC3.png
Attachment 9: PRM.png  332 kB  | Show | Hide all | Show all
Attachment 10: SRM.png  266 kB  Uploaded Tue May 3 15:11:05 2011  | Hide | Hide all | Show all
SRM.png
ELOG V3.1.3-