Figure.1 I-Q relative phase measurement as a function of LO power.
Blue curve : relative phase of AS55 that I have modified today (#4572).
Red curve : relative phase of AS11 that I had modified a week ago (#4554). Just for comparison.
The relative phase of AS55 agrees approximately what we expected according to the datasheet of PSCQ-2-51W. We expected 85 degree.

Figure.1 I-Q amplitude imbalance as a function of LO power.
From - 5 dBm to 5 dBm in LO power the imbalance is within 3 %.
But the precision of the measurement is also about 2 % (because I used an oscilloscope). Even so the imbalance is still good.
Quote from #4572 |
Some plots will be posted later.
|
|