Entry  Fri Apr 18 16:02:58 2008, alex, Update, SUS, end station sus front-end bug fix 
    Reply  Fri Apr 18 16:27:04 2008, rob, Update, SUS, end station sus front-end bug fix 
    Reply  Tue Apr 22 16:17:48 2008, rob, Update, SUS, end station sus front-end bug fix 
Message ID: 436     Entry time: Tue Apr 22 16:17:48 2008     In reply to: 425
Author: rob 
Type: Update 
Category: SUS 
Subject: end station sus front-end bug fix 

Quote:
installed and started new susEtmx.o and susEtmy.o to fix a problem with ETMY optical lever variables.


What Alex means is that the EPICS values for the ETMY optical levers were being clobbered in the RFM. The calculations were being done correctly in the FE, so the DAQ/testpoints were working--it was just the EPICS/RFM communication via c1losepics that was bugged. This was a result of the recent SUS code changes to accept inputs from the ASS for adaptive feedforward.