Entry  Tue May 18 11:54:29 2010, steve, Configuration, SAFETY, psl mechanical shutter is not working 
    Reply  Tue May 18 15:09:02 2010, steve, Configuration, SAFETY, PSL output shutter in order again 
Message ID: 2947     Entry time: Tue May 18 15:09:02 2010     In reply to: 2943
Author: steve 
Type: Configuration 
Category: SAFETY 
Subject: PSL output shutter in order again 

Quote:

As we learned yesterday, the PSL laser power out put mechanical shutter is not working in the remote mode. It only works in local manual mode.

Do not rely on the MEDM screen monitor readout! The position is only changing on the monitor. The main beam must be blocked before the output periscope.

Ben found the Sorenson 5V ps off. It was turned off since our last scheduled power outage. I wonder what else is running on 5V in the PSL? This power supply should be on the

"alarm handler"  list to avoid future repeat of this condition. However a real safety switch would have confirming position sensors of the shutter open or closed. Is there such thing at the sites?