Entry  Mon Nov 9 19:09:09 2009, Jenne, Update, PEM, Noise floor of the Ranger Seismometer Ranger_noiseFloor_Spectra.png
    Reply  Thu May 6 18:43:58 2010, rana, Update, PEM, Noise floor of the Ranger Seismometer ranger.pdf
Message ID: 2890     Entry time: Thu May 6 18:43:58 2010     In reply to: 2223
Author: rana 
Type: Update 
Category: PEM 
Subject: Noise floor of the Ranger Seismometer 

I added a noise model of the SR560 to the LISO opamp.lib. This assumes you're using it in G=100, low-noise mode. The voltage noise is correct, but I had to guess on the current noise because I didn't measure it before. Lame.

This can be compared with the noise that we measure when locking it down...

Attachment 1: ranger.pdf  34 kB  | Hide | Hide all