Thu Apr 8 10:17:52 2010, Alberto, Update, 40m Upgrading, REFL11 Noise Vs Photocurrent
|
Thu Apr 8 10:24:33 2010, Alberto, Update, 40m Upgrading, REFL11 Noise Vs Photocurrent
|
Thu Apr 8 20:53:13 2010, Koji, Update, 40m Upgrading, REFL11 Noise Vs Photocurrent
|
Fri Apr 9 06:45:28 2010, Alberto, Update, 40m Upgrading, REFL11 Noise Vs Photocurrent
|
Mon Apr 12 17:09:30 2010, Alberto, Update, 40m Upgrading, REFL11 Noise Vs Photocurrent
|
Mon Apr 12 22:44:30 2010, Koji, Update, 40m Upgrading, REFL11 Noise Vs Photocurrent
|
|
Message ID: 2783
Entry time: Thu Apr 8 10:24:33 2010
In reply to: 2782
|
Author: |
Alberto |
Type: |
Update |
Category: |
40m Upgrading |
Subject: |
REFL11 Noise Vs Photocurrent |
|
|
Quote: |
From the measurements of the 11 MHz RFPD at 11Mhz I estimated a transimpedance of about 750 Ohms. (See attached plot.)
|
Putting my hands ahead, I know I could have taken more measurements around the 3dB point, but the 40m needs the PDs soon. |