40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Thu Apr 1 00:43:33 2010, Mott, Update, General, PZT response for the innolight Innolight_Bode.png
    Reply  Thu Apr 1 07:17:15 2010, Koji, Update, General, PZT response for the innolight 
       Reply  Thu Apr 1 10:21:58 2010, Mott, Update, General, PZT response for the innolight Innolight_Response.pngLightwave_response.png
          Reply  Thu Apr 1 10:47:48 2010, Koji, Update, General, PZT response for the innolight 
             Reply  Thu Apr 1 12:07:22 2010, rana, Update, General, PZT response for the innolight 
                Reply  Thu Apr 1 18:05:29 2010, Mott, Update, General, PZT response for the innolight Innolight_Response_calFix.pngLightwave_response_calFix.png
                   Reply  Thu Apr 1 18:44:40 2010, Koji, Update, General, PZT response for the innolight 
                      Reply  Thu Apr 1 19:59:32 2010, Mott, Update, General, PZT response for the innolight Innolight_AM_Response.pngInnolight_AM_PM.pngInnoVsLW_PM.pngInnolight_AM_Response.pngLightwave_AM_PM.png
                         Reply  Mon Apr 12 14:20:10 2010, kiwamu, Update, Green Locking, PZT response for the innolight 
                            Reply  Tue Apr 13 19:53:06 2010, Mott, Update, Green Locking, PZT response for the innolight and lightwave 6x
Message ID: 2755     Entry time: Thu Apr 1 18:44:40 2010     In reply to: 2754     Reply to this: 2756
Author: Koji 
Type: Update 
Category: General 
Subject: PZT response for the innolight 

Innolight 10 rad/V @ 100kHz => 1e6/f rad/V => 1MHz/V

LWE 30 rad/V @ 100kHz => 3e6/f rad/V => 3MHz/V

---------

BTW, don't let me calculate the actuator response everytime.

The elog (=report) should be somewhat composed by the following sections

Motivation - Method - Result (raw results) - Discussion (of the results)

Quote:

  We realized that we had measured the wrong calibration value; we were using the free-running error signal with the marconi far from the beat frequency, which was very small.  When we put the Marconi right at the beat, the signal increased by a factor of ~12 (turning our original calibration of 10 mV/rad into 120 mV/rad).  The re-calibrated plots are attached. 

 