Entry  Wed Jan 20 20:11:56 2010, Alberto, Update, ABSL, Some preliminary results from measuring PRC's transmissivity for an amplitude modulated beam 2010-01-19_PRCtransmissivityVsModel.png
    Reply  Thu Jan 21 10:31:13 2010, Koji, Update, ABSL, Some preliminary results from measuring PRC's transmissivity for an amplitude modulated beam 
Message ID: 2536     Entry time: Thu Jan 21 10:31:13 2010     In reply to: 2534
Author: Koji 
Type: Update 
Category: ABSL 
Subject: Some preliminary results from measuring PRC's transmissivity for an amplitude modulated beam 

Nice and interesting plot.

I suppose slight decrease of the Schnupp asymmetry (in your model) adjusts the discrepancy in the high freq region.
At the same time, it will make the resonance narrower. So you need to put some loss at the recombination (=on the BS)?

...of course these depends on the flatness of the calibration.