Entry  Fri Dec 11 08:35:04 2009, steve, Configuration, VAC, pumpdown has started wiping.jpg
    Reply  Fri Dec 11 13:18:17 2009, Alberto, Configuration, VAC, pumpdown has started 
       Reply  Fri Dec 11 14:19:22 2009, Koji, Configuration, VAC, pumpdown has started 
    Reply  Fri Dec 11 15:21:40 2009, steve, Configuration, VAC, pumpdown#67 is completed pd7h.jpgoplevpd8h.jpg
Message ID: 2400     Entry time: Fri Dec 11 15:21:40 2009     In reply to: 2394
Author: steve 
Type: Configuration 
Category: VAC 
Subject: pumpdown#67 is completed 

Quote:

Oplev positions before and after drag wiped arm TMs as of yesterday. Slow-mode pumpdown has started with 3/4 turn opened  RV1 valve at 8am today.

 Pump down is completed. Valve configuration is VACUUM NORMAL. CC1 pressure is in the ~8 e-5 torr   PSL output shutter is opened.

Attachment 1: pd7h.jpg  65 kB  | Hide | Hide all
pd7h.jpg
Attachment 2: oplevpd8h.jpg  105 kB  | Hide | Hide all