Attached is the schematic of the Piezo Jena driver measuring box made in a Pomona box:
2.2 uF
In ----o-------- | | --------o-------- Out
| |
_ |
_ 1uF R 7.5 kOhms
| |
| |
GND GND
The 1 uF cap is there to simulate the piezo and the 2.2 uF and 7.5k resistor ac couple the signal for the spectrum analyzer. They give a ~10 Hz corner frequency. |
|
|
|
|