Message ID: 2023     Entry time: Tue Sep 29 22:51:20 2009     In reply to: 2020
Author: Koji 
Type: Update 
Category: MZ 
Subject: Possible gain mis-calibration at other places (Re: MZ work done) 

Probably there is the same mistake for the PMC gain slider. Possibly on the FSS slider, too???

Quote:

2) The EPICS setting for the MZ gain slider was totaly wrong.
    Today I learned from the circuit, the full scale of the gain slider C1:PSL-MZ_GAIN gave us +/-10V at the DAC.
    This yield +/-1V to V_ctrl of the AD602 after the internal 1/10 attenuation stage.
    This +/-1V didn't correspond to -10dB~+30dB, but does -22dB~+42dB and is beyond the spec of the chip.