Message ID: 1953     Entry time: Wed Aug 26 16:35:03 2009     In reply to: 1950
Author: Alberto 
Type: Configuration 
Category: PSL 
Subject: PSL reference cavity temperature box modifications 

Basically, in addition to the replacement of the resistors with metal film ones, Peter replaced the chip that provides a voltage reference.

The old one provided about 2.5 V, whereas the new one gets to about 7V. Such reference voltage somehow depends on the room temperature and it is used to generate an error signal for the temperature of the reference cavity.

Peter said that the new higher reference should work better.