Entry  Tue Jul 19 18:41:42 2022, yuta, Update, BHD, Contrast measurements for Michelson and ITM-LO ContrastMeasurements.pdf
    Reply  Wed Jul 20 18:07:52 2022, Paco, Update, BHD, BHD MICH test BHD_MICH_OSC.pdf
       Reply  Tue Jul 26 20:54:08 2022, Paco, Update, BHD, BHD MICH test - LO phase control 20220726_BICHD.pdf
          Reply  Wed Jul 27 18:30:50 2022, yuta, Update, BHD, LO beam power at BHD DCPDs is significantly lower than expected LOBeamAtBHD.JPGITMXSingleBounceAtBHD.jpg
             Reply  Thu Jul 28 14:34:40 2022, Yehonathan, Update, BHD, LO beam power at BHD DCPDs is significantly lower than expected 
                Reply  Thu Jul 28 16:51:55 2022, Tega, Update, BHD, Shaking test for LO beam AS beam to BHD DCPDs Screenshot_2022-07-28_16-58-34_LOandASShaking.pngScreenshot_2022-07-28_18-08-55_DCPDPOPSuspensionCoherence.png
                   Reply  Thu Jul 28 20:16:26 2022, Anchal, Update, BHD, Shaking test for LO beam AS beam to BHD DCPDs 
                   Reply  Fri Jul 29 18:24:53 2022, Tega, Update, BHD, LO beam power improved by factor of 6 after LO and AS beam alignment LO_and_oldAS_settings.pngBHD_fringe_settings.pngScreenshot_2022-07-29_19-37-39.pngFromTheLeft-AS-POP-LO.JPG
                      Reply  Wed Aug 3 16:00:51 2022, yuta, Update, BHD, BHD fringe aligned with reduced LO and AS beam clipping Screenshot_2022-08-03_15-46-36_BHDfringeAlmostUnclipped.png
                         Reply  Tue Aug 9 15:33:12 2022, yuta, Update, BHD, BHD fringe contrast improved from 43% to 74% HPC-DCPD_B_OUT_1344118517_ITMY-LO.pngHPC-DCPD_A_OUT_1344118517_ITMY-LO.pngHPC-DCPD_B_OUT_1344118318_ITMX-LO.pngHPC-DCPD_A_OUT_1344118318_ITMX-LO.png
                            Reply  Tue Aug 9 15:50:22 2022, Koji, Update, BHD, BHD fringe contrast improved from 43% to 74% 
                               Reply  Wed Nov 16 15:09:08 2022, yuta, Update, BHD, BHD fringe contrast measured with unwhitening filters ContrastMeasurements20221116_edited.pdf
                         Reply  Tue Oct 25 08:25:00 2022, JC, Update, BHD, BHD fringe aligned with reduced LO and AS beam clipping 
                         Reply  Thu Nov 17 11:24:39 2022, JC, HowTo, LSC, Locking MICH 
                            Reply  Tue Nov 29 11:38:37 2022, JC, HowTo, LSC, Lock Single Arm After MICH lock Screen_Shot_2022-11-29_at_11.01.56_AM.png
Message ID: 17045     Entry time: Thu Jul 28 20:16:26 2022     In reply to: 17044
Author: Anchal 
Type: Update 
Category: BHD 
Subject: Shaking test for LO beam AS beam to BHD DCPDs 

Some insights from the inside vacuum situation:

  • The beam is an incident near normal on PR2 close to the center of the optic. It wasn't hard to align this part, I'm very confident that we aligned it to the center of PR2. So I do not think the LO beam is ghost beam from PR2.
  • The place that is most susceptible to clipping is POP_SM5 mirror in front of LO1. The LO beam has little clearance from the edge of the mirror.
  • Another possibility of clipping in LO beam is through the cage of LO2. LO2 is a 45-degree incidence mirror, so it is possible we are clipping off the cage or seeing a ghost beam mixed in LO beam here.
  • The fact that moving PR2 is affecting LO beam is weird but doesn't necessarily mean it is a ghost from PR2.