Message ID: 17035     Entry time: Mon Jul 25 18:22:30 2022
Author: Deeksha 
Type: Summary 
Category: Wiki 
Subject: Measured the PZT TF Successfully  

Measured the PZT beatnote using the setup mentioned in elog post 17031. Attached is the data taken from 10kHz to 1MHz, decadewise data was also taken that I'm not including in this post. A_R refers to the transfer function taken of channel A wrt the voltage reference (the swept sine we are inputting which has an IF of 30kHz). A and B correspond to the I and Q components of the signal taken from the DFD, respectively. I am currently working on plotting the data, and will shortly update this post with plots. Next steps - 

- quantify the uncertainty in the signal (I think)

- vectfit the data to find poles and zeroes

(and possibly find a better way to print/obtain data)

Edit: first pass of data plotted

Attachment 1: A_R_MAG.txt  56 kB  Uploaded Mon Jul 25 19:33:30 2022  | Hide | Hide all
"4395A REV1.12"
"DATE: Sep 17 2017"



"CHANNEL: 1"
"MEASURE TYPE: A/R"
"FORMAT TYPE: LOG MAG"
"NUMBER of POINTS: 801"
"SWEEP TIME:  385.3 ms"
... 811 more lines ...
Attachment 2: A_R_PHASE.txt  56 kB  Uploaded Mon Jul 25 19:33:36 2022  | Hide | Hide all




"CHANNEL: 2"
"MEASURE TYPE: A/R"
"FORMAT TYPE: PHASE (DEG)"
"NUMBER of POINTS: 801"
"SWEEP TIME:  385.3 ms"
... 808 more lines ...
Attachment 3: B_R_MAG.txt  56 kB  Uploaded Mon Jul 25 19:33:44 2022  | Hide | Hide all
"4395A REV1.12"
"DATE: Sep 17 2017"



"CHANNEL: 1"
"MEASURE TYPE: B/R"
"FORMAT TYPE: LOG MAG"
"NUMBER of POINTS: 801"
"SWEEP TIME:  385.3 ms"
... 809 more lines ...
Attachment 4: B_R_PHASE.txt  56 kB  Uploaded Mon Jul 25 19:33:53 2022  | Hide | Hide all




"CHANNEL: 2"
"MEASURE TYPE: B/R"
"FORMAT TYPE: PHASE (DEG)"
"NUMBER of POINTS: 801"
"SWEEP TIME:  385.3 ms"
... 807 more lines ...
Attachment 5: freq_resp_I.png  144 kB  Uploaded Tue Jul 26 11:21:39 2022  | Hide | Hide all
freq_resp_I.png
Attachment 6: freq_resp_Q.png  150 kB  Uploaded Tue Jul 26 11:21:59 2022  | Hide | Hide all
freq_resp_Q.png