Message ID: 17008     Entry time: Fri Jul 15 22:36:04 2022     In reply to: 17007
Author: rana 
Type: Summary 
Category: LSC 
Subject: FPMI with REFL/AS55 demod phase adjust 

Very nice!

DARM feedback should go to ETMY - ETMX, not just a single mirror: Differential ARM.

For it to work with 1 mirror the UGF of the CARM loop must be much larger than DARM UGF. But in our case, both have a UGF of ~150 Hz.

In principle, you could run the CARM loop with higher gain by using the CM servo board, but maybe that can wait until the X,Y -> CARM, DARM handoff.

 