To give more alignment ranges for the SUS alignment, we started updating the output resistors of the BHD SUS coil drivers.
As Paco has already started working on LO1 alignment, we urgently updated the output Rs for LO1 coil drivers.
LO1 Coil Driver 1 now has R=100 // 1.2k ~ 92Ohm for CH1/2/3, and LO1 Coil Driver 2 has the same mod only for CH3. JC has taken the photos and will upload/update an elog/DCC.

We are still working on the update for the SR2, LO2, AS1, and AS4 coil drivers. They are spread over the workbench right now. Please leave them as they're for a while.
JC is going to continue to work on them tomorrow, and then we'll bring them back to the rack.

Attachment 1: IMG_0527.jpeg  2.671 MB  Uploaded Thu Apr 21 08:19:20 2022  | Hide | Hide all
IMG_0527.jpeg
Attachment 2: IMG_0528.jpeg  619 kB  Uploaded Thu Apr 21 08:19:34 2022  | Hide | Hide all
IMG_0528.jpeg
Attachment 3: IMG_0529.jpeg  611 kB  Uploaded Thu Apr 21 08:19:44 2022  | Hide | Hide all
IMG_0529.jpeg
Attachment 4: IMG_0530.jpeg  2.364 MB  Uploaded Thu Apr 21 08:19:53 2022  | Hide | Hide all
IMG_0530.jpeg
Attachment 5: IMG_0531.jpeg  617 kB  Uploaded Thu Apr 21 08:20:06 2022  | Hide | Hide all
IMG_0531.jpeg
Attachment 6: IMG_0532.jpeg  431 kB  Uploaded Thu Apr 21 08:20:11 2022  | Hide | Hide all
IMG_0532.jpeg
ELOG V3.1.3-