[Koji, Ian]
1) We attached the 30 coil driving cables to the vacuum feed through to the sat amp [40m wiki] they run along the cable tray then up and down into the rack.
2) we checked all DB and power cables. We found that the anti-imaging filter had a short and got very hot when plugged in. the back power indicator lights turned on fine but the front panel stayed off. We removed it and replaced it with the one that was on the test stand marked for the BHD. This means we need to fix the broken one and Koji mentioned getting another one.
3) we reassigned the ADC and DAC channels in the iscey model and the asy model. we committed a version before we made any changes.
4) Finally we tested the setup to make sure the ETM was being damped.
Next step:
1) Measure the change of the PD gains and the actuator gains. See pervious elog |